598209
17
Zoom out
Zoom in
Previous page
1/20
Next page
Chapter II
Portable digital multimeter 29
5.6. Capacitance
)
Note Discharge all capacitors before taking measurements.
Ranges Accuracy
Measurement
current
Max measurement
time
Protection
Resolution
50 nF∗∗
100 nA 10 pF
500 nF 1 µA 100 pF
5 µF 10 µA 1 nF
50 µF 100 µA
0.5 s
10 nF
500 µF
1% R + 2D
1.5 s 100 nF
5000µF 3 s/mF 1 µF
50 mF
2% R + 2D
1 mA
3 s/mF
600 V
RMS
10 µF
Overload protection can be reset automatically.
∗∗ It is highly recommended to use very short and shielded test leads for measurements in
this range.
Number of count : 5000
Range selection : automatic or manual
Maximum open circuit voltage : 7 V
5.7. Diode threshold voltage measurement
Measurable voltages : 0 to 2 V
Measurement current : 1 mA typical
Resolution : 1 mV
Protection : 600 V
RMS, can be reset automatically
5.8. Frequencies
Selector switch setting : V
AC
+ SEL/ON
Measurement range : 0.62 Hz to 500 kHz
Accuracy : 0.03 %
Protection : 600 V
RMS
Display : 50.000 count
Range Sensitivity
5 V to 500 V 600 V
0.62 Hz to 5 kHz
2 % of range 100 V
5 kHz to 50 kHz 5 % of range 250 V
50 kHz to 500 kHz 10 % of range ////////////////
Rectangular signals
5.9. Duty cycle %+, %-
θ
T
θ
T
% - =
θ
T
x 100
% + =
θ
T
x 100
Resolution : 0.01 %
Min. duration for θ or T-θ : 2 µs
Max. duration for T : 0.8 s
Min. duration for T : 100 µs
Absolute error as a % :
T
4
10.3
Sensitivity
: see frequency measurement mode
Absolute additional error on θ, due to the zero crossing slope :
0.1 x
P
C
with C : r
ange in V (C = 5000 V for 600V
DC
or 600 V
AC
ranges)
with P : slope in V/s
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrix MX 58HD at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrix MX 58HD in the language / languages: English as an attachment in your email.

The manual is 0,44 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrix MX 58HD

Metrix MX 58HD User Manual - German - 20 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info