598209
16
Zoom out
Zoom in
Previous page
1/20
Next page
Chapter II
28 Portable digital multimeter
5.3. DC current
Selector
switch
Accuracy
Max. voltage
drop
Protection
Fuses
Resolution
5 mA 1 µA
50 mA
700 mV
10 µA
500 mA
0.2 % R + 2 D
1.5 V
F1 + F2
100 µA
10 A∗∗
0.5 % R + 5 D 500 mV
600 V
RMS
F2 10 mA
refer to fuse specifications section 6.1.1.
∗∗ acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests
Number of count : 5000
Additional error in Pk +/- mode for a pulse width of 1 ms : 1 % R ± 50 D
For measurements performed on alternative signals, the selected range must tally with the max.
value of the signal peak.
5.4. AC currents (AC and AC + DC)
Accuracy
40 Hz to
5 kHz
5 kHz to
20 kHz
20 kHz
to 30kHz
Ranges
5 % to 100 % of range
Protect.
Fuses
Resol.
Max.
Crest
Max.
voltage
drop
5 mA 1%R+3D 1.5% typ. 4% typ. 1 µA 10 mA
50 mA 1.5% typ. 4% typ. 10 µA 100 mA
700 mV
500 mA
1%R +3 D
1.5% typ. 4% typ.
F1 + F2
100 µA 1 A 1.5 V
10 A∗∗
1.5%R+3 D
Æ 2 kHz
1.5% typ. ///////////
600
V
RMS
F2 10 mA 30 A 500 mV
refer to fuse specifications section 6.1.1.
∗∗ acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests
Specification : from 5 % to 100 % of the range
Number of count : 5000
Additional error according to crest factor : 0 % for a crest factor < 1.5
1 % for a crest factor of 1.5 to 2
4 % for a crest factor of 2 to 3
Additional error in I
AC+DC with a direct current at input : 1 %
5.5. Resistance / Continuity
Ranges Accuracy
Measur. current
Protection
Resolution
500 Ω/
0.1% R + 5 D
1 mA
100 mΩ
5 kΩ
100 µA
1 Ω
50 kΩ
10 µA
10 Ω
500 kΩ
0.1% R + 3 D
1 µA
100 Ω
5 MΩ∗∗
0.3% R + 3 D 100 nA
1k Ω
50 MΩ∗∗
1% R + 3 D 50 nA
600 V
RMS
10 k Ω
Overload protection can be reset automatically.
∗∗ The use of very short and shielded test leads is highly recommended for measurements in
this range (> 1 MΩ).
Number of count : 5000
Range selection : automatic or manual (set in continuity mode)
Maximum open circuit voltage : 7 V
Detection threshold in continuity mode : 10 Ω to 20 Ω
Response time in continuity mode : 1 ms
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrix MX 58HD at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrix MX 58HD in the language / languages: English as an attachment in your email.

The manual is 0,44 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrix MX 58HD

Metrix MX 58HD User Manual - German - 20 pages


The manual is sent by email. Check your email