MI 3100 s, MI 3100 SE EurotestEASI Measurements
43
5.4.4 RCD Auto-test
RCD auto-test function is intended to perform a complete RCD test (trip-out time at different
residual currents, trip-out current and contact voltage) in one set of automatic tests, guided by
the instrument.
Additional key:
HELP Click Toggles between top and bottom part of results
field.
Keep pressed for 1s Enters Help screen
RCD Auto-test procedure
RCD Auto-test steps Notes
Select the RCD function using the function selector
switch.
Set sub-function to AUTO using UP / DOWN keys.
Set test parameters (if necessary).
Connect test cable to the instrument.
Connect test leads to the item to be tested (see Figure
5.17).
Press the TEST key to perform the test. Start of test
Test with IN, 0 (step 1). RCD should trip-out
Re-activate RCD.
Test with IN, 180 (step 2). RCD should trip-out
Re-activate RCD.
Test with 5IN, 0 (step 3). RCD should trip-out
Re-activate RCD.
Test with 5IN, 180 (step 4). RCD should trip-out
Re-activate RCD.
Test with ½IN, 0 (step 5). RCD should not trip-
out
Test with ½IN, 180 (step 6). RCD should not trip-
out
Trip-out current test, 0 (step 7). RCD should trip-out
Re-activate RCD.
Trip-out current test, 180 (step 8). RCD should trip-out
Re-activate RCD.
Store the result by pressing the MEM key
(MI 3100 SE only).
End of test
Result examples:
Step 1
Step 2
43


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI3100S at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI3100S in the language / languages: English as an attachment in your email.

The manual is 1.81 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI3100S

Metrel MI3100S User Manual - German - 92 pages

Metrel MI3100S User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info