MI 3121H Smartec 2,5 kV Insulation / Continuity Continuity
29
5.2.3 Compensation of test leads resistance
This chapter describes how to compensate for test leads resistance in both continuity
functions (RLOW and CONT). Compensation is required to eliminate the influence of test
leads resistance and the internal resistances of the instrument on the measured
resistance. The lead compensation is therefore a very important feature to obtain correct
result. Once compensation has been performed, the compensation symbol ( ) appears
on the screen.
Each of RLOW and CONT has each own compensation.
Circuits for compensating the resistance of test leads
Figure 5.16: Shorted test leads
Compensation of test leads resistance procedure
Select the any continuity function using the function switch.
Connect test lead to the top of the instrument and short the test leads together
(see Figure 5.16).
Press TEST to perform resistance measurement.
Press the CAL key to compensate leads resistance.
Note:
The limit value for lead compensation is 5 Ω.
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3121H SMARTEC 2.5 kV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3121H SMARTEC 2.5 kV in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3121H SMARTEC 2.5 kV

Metrel MI 3121H SMARTEC 2.5 kV User Manual - German - 44 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info