MI 2092 Power Harmonics Analyser Section VI Theory of operation
57
Example 2:
The same set-up as above but:
Statistical Analysis is enabled
Phase 1 voltage harmonics (as many as possible) are to be logged.
The Instrument can calculate a Statistical Analysis for all recorded parameters except
current and voltage harmonics; thus 12 channels are required for recording statistics.
A maximum of 64 channels can be selected for data logging. The settings for
Example 1 above require 12 channels, leaving 48 channels free. The Instrument can
record harmonics up to 41
st
, requiring 40 channels (all possible harmonics from the 2
nd
to the 41
st
will be recorded). Thus a total 52 channels need to be set for Periodic
Analysis.
The number of records that the Instrument can store with this setting is given by:
Rn
max
= (2032-12) *1024 / (12 + 52 * 6) = > 6384
(Record Length = 12 + 52 * 6 = 324 bytes)
Thus, if recording is required over one week (604,800 seconds), the shortest recording
interval is given by:
Interval = 604,800 seconds / 6384
94.7seconds
Thus setting an IP of 2 minutes will produce 5040 record per week in memory. This will
leave the following memory free:
FREE Memory = Memory Capacity - Memory Used
(2032 - 12) * 1024 - 5040 * 324
435,520 bytes.
This would allow recording over a further 44 hours (with no Voltage Anomalies), or for
recording 2655 Voltage Anomalies.
57


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 2092 Power Harmonics Analyser at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 2092 Power Harmonics Analyser in the language / languages: English as an attachment in your email.

The manual is 1,58 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 2092 Power Harmonics Analyser

Metrel MI 2092 Power Harmonics Analyser User Manual - Dutch - 46 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info