pump will assess channel changing needs based on "noise" levels detected during
an energy scan. The pump will perform the energy scan if after 10 minutes no
CGM transmitter signal has been received. If the channel change occurs the pump
will send beacons on the new channel.
The CGM transmitter will initiate a channel search when beacon detection fails on
the associated channel. The search will be conducted across all five channels.
When the beacon is located the transmitter will rejoin on the identified channel.
Upon re-association any missed packets (up to 10 hours) will be transmitted from
the CGM transmitter to the pump.
In normal operation the CGM transmitter will transmit a packet every 5 minutes
and retransmit the packet if the data is corrupted or missed.
Radio frequency (RF) communications specifications
Utilizes the IEEE 802.15.4 protocol with the proprietary data format.
Pump frequency 2.4 GHz; proprietary Medtronic protocol; range up to
1.8 meters (6 feet)
Maximum output
power (EIRP)
-4 dBm (.398 mW)
Operating frequencies 2420 MHz, 2435 MHz, 2450 MHz, 2465 MHz, 2480 MHz
Bandwidth 5 MHz which is allocated channel bandwidth per the IEEE
protocol
Data security
The MiniMed 640G insulin pump is designed to only accept radio frequency (RF)
communications from recognized and linked devices (you must program your
pump to accept information from a specific device).
The MiniMed 640G system ensures data security via proprietary means and ensures
data integrity using error checking processes, such as cyclic redundancy checks.
Product specifications and safety information
Product specifications and safety information 269
283


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medtronic MiniMed 640G at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medtronic MiniMed 640G in the language / languages: English as an attachment in your email.

The manual is 23.36 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Medtronic MiniMed 640G

Medtronic MiniMed 640G User Manual - German - 365 pages

Medtronic MiniMed 640G Additional guide - Dutch - 1 pages

Medtronic MiniMed 640G User Manual - Dutch - 331 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info