6 Miscellaneous
GB
6.3
Disposal
The air hose is not connected or leaking. Check the connection between the
upper arm cuff and the unit and check the air hose for leaks. Repeat the
measurement.
“EE” appears in the display
The unit has not been inflated enough. Repeat the measurement
A problem occurred during inflation or during the measurement. Repeat the
measurement.
No display or display locked
• No batteries have been inserted, batteries wrongly inserted or batteries are
completely empty. Insert batteries, check battery position or replace with
new batteries.
• A malfunction or fault has occurred. Briefly remove the batteries and then
replace them.
Please contact the service centre if the error message appears again
when the measurement is repeated.
Remove the batteries before cleaning the unit.
Never use strong detergents or hard brushes.
Clean the unit using a soft cloth lightly moistened with a mild soapy
solution. Never use abrasive cleaning agents or alcohol.
• Do not allow any water to get into the unit. Do not use the unit again until
it is completely dry.
Remove the batteries from the unit if you do not wish to use it for an
extended period of time, otherwise there is a risk of battery leakage.
Do not expose the unit to direct sunlight, and protect it from dust and
moisture.
Only inflate the cuff when it is in position around the upper arm.
The unit is best kept in the storage pouch included in your purchase and
should preferably be stored in a clean dry place.
Servicing and calibration (applies to commercial use):
The unit has been calibrated by the manufacturer for a period of two years’
use. In the case of commercial use, the unit must be serviced and recali-
brated at least once every two years. This calibration will be charged for and
can be carried out by an appropriate authority or an authorised service
centre - in accordance with the regulations for users of medical equipment.
This product must not be disposed together with the domestic waste.
All users are obliged to hand in all electrical or electronic devices, regardless
of whether or not they contain toxic substances, at a municipal or commer-
cial collection point so that they can be disposed of in an environmentally
acceptable manner.
Please remove the batteries before disposing of the equipment. Do not dispo-
se of old batteries with your household waste, but dispose of them at a battery
collection station at a recycling site or in a shop.
Consult your municipal authority or your dealer for information about disposal.
6.2
Care and
maintenance
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Medisana MTV 51120 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Medisana MTV 51120 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Swedish, Spanish, Finnish as an attachment in your email.

The manual is 1,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info