Front Panel Operation
3-34 2001-900-01 Rev. K/ August 2010
A procedure to measure in-circuit current follows:
1. Select the in-circuit current measurement mode from the
CONFIGURE DCI menu and place the instrument in
the DCI function. Note that the 4W annunciator lights to
indicate this is a 4-wire measurement.
2. Connect a set of Kelvin test probes, such as Keithley
Model 5805 or 5806, to the Model 2001 INPUT HI and
LO terminals and SENSE HI and LO terminals.
3. For R
TRACE
<50m, or where the conductors are physi-
cally hot, rel out (zero correct) any thermal offsets that
are present before measuring in-circuit current. With
power to the test circuit removed, place the probes on
the desired trace and enable rel from the DCI function.
4. Turn on the power to the test circuit and read the in-cir-
cuit current calculation. The speed of this measurement
is four readings per second at 1PLC. Its range is fixed at
12A. The default filter for in-circuit current is a moving
average of ten readings. Additional filtering may be
needed at low voltage levels.
COUPLING
This parameter selects the input coupling for the ACI
function. When AC coupling is selected, a DC blocking
capacitor is placed in series with the AC measurement
circuit. (Note that the current shunt resistors are always DC
coupled to the inputs.) This removes the DC component
from the RMS and average ACI measurement.
When AC+DC coupling is selected, the blocking capacitor is
removed. Subsequent RMS or average ACI measurements
will reflect both the AC and DC components of the signal.
You can set the ACI coupling as follows:
1. From the CONFIGURE ACI menu, select COUPLING
and press ENTER. The following menu is displayed:
SET AC COUPLING
AC AC+DC
2. Highlight the desired coupling and press ENTER.
Note that the coupling settings for ACI and ACV are
discrete. Thus, setting the coupling of ACI has no effect on
the coupling of ACV.
AC-TYPE
This parameter selects the measurement type for the ACI
function. The Model 2001 directly measures RMS and
average AC current. You can program the ACI type
parameter as follows:
1. From the CONFIGURE ACI menu, select AC-TYPE
and press ENTER. The following menu is displayed:
SET ACI MEASUREMENT
RMS AVERAGE
2. Highlight the desired measurement type and press EN-
TER.
With this parameter selected, the instrument performs
RMS AC current measurements.
When this item is selected, the signal path in
the instrument bypasses the RMS converter, so the average
ACI measurement is just the filtered output of a full wave
rectifier.
Multiple displays
The multiple displays for AC current that show multiple
functions are shown in Figure 3-13. There are no
multifunction displays for DC current.
66


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2001 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2001 in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info