598220
65
Zoom out
Zoom in
Previous page
1/476
Next page
Front Panel Operation
2001-900-01 Rev. K/ August 2010 3-33
2. Highlight the desired mode and press ENTER.
NORMAL: This option is for normal current measuring,
where the meter is placed in series with the current path and
the voltage across an internal shunt resistor is measured. The
Specifications are derated by 50 ppm for currents over 0.5 A
because of the self-heating effects on the shunt resistor.
IN-CIRCUIT: In-circuit current is a calculation based on a
4-wire resistance measurement and a voltage measurement.
It is similar to an offset-compensated ohms reading.
A measurement overflow occurs for any of the following
conditions:
The measured voltage exceeds |±200 mV|.
The trace resistance is less than 1m or greater than
10.
The in-circuit current is greater than 12 A.
The current in a low resistance conductor (e.g., a printed
circuit trace) can be measured without breaking the current
path. The Model 2001 can do this with a pair of Kelvin test
probes across the conductor. See Figure 3-12. The method
follows:
1. Using one set of the Kelvin probe tips, the instrument
sources a known current (I
SOURCE
) through the conductor
and simultaneously measures the resulting voltage
(V
MEAS1
) with the other set of probe tips:
or
2. The instrument then measures the voltage (V
MEAS2
)
across the conductor without sourcing an additional cur-
rent:
or
3. It then calculates the in-circuit current by combining the
equations and solving for I
IN-CKT
:
Because of accuracy considerations, in-circuit current
readings are limited to traces with a resistance of 1 m to
10 . If either of these limits is exceeded in the resistance
calculation, the in-circuit current cannot be calculated. The
bottom line of the front panel display will show and update
the trace resistance.
cáÖìêÉ=PJNO
DC in-circuit current measurements
NEXT
DISPLAY
PREV
POWER
DCV ACV DCI ACI Ω2 Ω4
FREQ TEMP
REL TRIG STORE RECALL
INFO LOCAL CHAN SCAN CONFIG MENU EXIT ENTER
RANGE
AUTO
FILTER MATH
RANGE
2001 MULTIMETER
SENSE
Ω 4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
Model 2001
Circuit
Under Test
Caution : Maximum Input = +200mV on (I
IN - CKT
+ I
SOURCE
) • R
TRACE
,
where I
SOURCE
= 10mA.
+00.207 ADC ICkt
Trace resistance: 1.0000Ω
Y
Y
x
Note: The distance
"X" must be greater
than 10 times the
distance "Y" or
additional errors will
be introduced.
V
MEAS1
I
IN-CKT
I
SOURCE
+R
TRACE
=
R
TRACE
V
MEAS1
I
IN-CKT
I
SOURCE
+
-------------------------------------------------=
V
MEAS2
I
IN-CKT
R
TRACE
=
R
TRACE
V
MEAS2
I
IN-CKT

----------------------=
V
MEAS1
I
IN-CKT
I
SOURCE
+
-------------------------------------------------
V
MEAS2
I
IN-CKT

----------------------=
V
MEAS1
I
IN-CKT
V
MEAS2
I
IN-CKT
I
SOURCE
+=
V
MEAS1
I
IN-CKT
-V
MEAS2
I
IN-CKT
V
MEAS2
I
SOURCE
=
I
IN-CKT
V
MEAS1
-V
MEAS2
=V
MEAS2
I
SOURCE

I
IN-CKT
V
MEAS2
I
SOURCE

V
MEAS1
-V
MEAS2

---------------------------------------------=
65


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Keithley 2001 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Keithley 2001 in the language / languages: English as an attachment in your email.

The manual is 2,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info