Care and Cleaning
1. The probe (sensor) is the most intricate part of the thermometer and should be kept
clean and intact to acquire accurate readings.
Use the following method to clean the probe:
1) Gently swab the surface of the probe using a cotton bud soaked in > 95%
medical alcohol.
2) Allow at least 1 minute for the probe to fully dry.
2. If the probe (sensor) is broken, please contact customer services.
3. Use a piece of soft, dry cloth to clean the display screen and external surface of the
thermometer. If the thermometer is very dirty, the cloth can be moistened with some
medical alcohol to clean the device.
4. The product is not waterproof. Do not clean the device with detergents. Do not soak
the thermometer in water or other liquids.
Maintenance
1. This company has not authorized any agency or individual to carry out product
repairs or maintenance. Do not attempt to disassemble or modify the thermometer if
you suspect functional issues with the device.
2. The IR thermometer is an extremely precise instrument. Any improper maintenance,
disassembly, or modification may lead to inaccuracies of the product measurements.
3. Please check the device if damaged after it is dropped. If unsure, please contact
customer services for having the device checked.
4. The thermometer is initially calibrated at the time of manufacture. If this thermometer
is used according to the use instruction, periodic re-adjustment is not required. If any
time your question the accuracy of measurement during the warranty period, please
contact customer services.
5. No component can be maintained by user in the monitor. circuit diagrams,
component part lists, descriptions, calibration instructions, or other information which
will assist the user’s appropriately qualified technical personnel to repair those parts
of equipment which are designated repairable can be supplied.
6. The method for verifying the clinical accuracy can be requested, please contact
customer services.
7. The device must not be stored or used at an excessively high or low temperature or
humidity (see technical data), in sunlight, in association with an electrical current or
in dusty locations. Avoid dropping or subjecting the product to external forces.
Otherwise inaccuracies can occur.
8. Do not directly touch the probe with your fingers or blow on it. Measurements taken
using a damaged or dirty IR probe may be inaccurate.
9. The monitor requires 4 hours to warm from the minimum storage temperature between
uses until the monitor is ready for its INTENDED USE when the ambient temperature is
20 °C (68°F).
10. The monitor requires 4 hours to cool from the maximum storage temperature between
uses until the monitor is ready for its INTENDED USE when the ambient temperature is
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for iHealth PT2L at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of iHealth PT2L in the language / languages: English as an attachment in your email.

The manual is 0.46 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of iHealth PT2L

iHealth PT2L User Manual - German - 16 pages

iHealth PT2L User Manual - French - 15 pages

iHealth PT2L User Manual - Spanish - 15 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info