21 Appendix
21.1 Evaluation of Measured Values for Individual Measurements
as well as for Calculated Quantities
In order to assure that the limit values for the individual measurements
are always observed, device measuring error must be taken into
consideration.
The table in the appendix allows for calculation of the required minimum
display value for each respective measurement which must appear at the
instrument in consideration of measuring error (under nominal conditions
of use), in order to assure that the required limit value is not fallen short of
(DIN VDE 0413, part 1). Intermediate values can be interpolated.
Measuring Error for Test Sequences
The test instrument takes respective measuring error into consideration
during automatic test sequences, and corrected results are entered into
the test report, as long as this function has been activated in the setup
menu under “include service error.
Omitting the Protective Conductor Test in the Case of Fully Insulated Devices
You are testing a fully insulated safety class I device (e. g. screen, sub-
mersible pump, etc.), which is not equipped with an external protective
conductor contact.
The decision as to the necessity of a protective conductor test in this case is
to be taken by a qualified electrician who should also assume responsibility.
You can omit the protective conductor test by pressing the key as soon
as the following instruction is shown: „Please connect the probe with the
protective conductor of the DUT“.
Tables for the calculation of minimum display values for insulation
resistance and maximum display values for protective conductor resis-
tance, equivalent leakage current, probe current and residual current in
consideration of device measuring error:
1)
Depending upon resolution
21.2 Evaluation of Measured Values during Equivalent Leakage Current
Measurement (Automatic Test Sequence According to Standard)
During equivalent leakage current measurement, L and N are intercon-
nected whereupon a test voltage of 230 V is applied between LN and PE
and the leakage current is measured. Thus the most adverse case (N in-
terrupted) is being tested.
As a rule, this results in at least double the value of the direct leakage cur-
rent measurement (since all discharge capacitors are placed in parallel).
If frequency converters are used in addition to that, it is no longer possible
to compare the measured values between direct leakage current and
equivalent leakage current method. In this case, we recommend conduct-
ing individual measurements in accordance with the residual current
method.
R
INS
M R
PE
Limit Value
Minimum
Display Value
Limit Value
Maximum
Display Value
0.100 0.115 0.100 0.085
0.250 0.273 0.200 0.180
0.500 0.535 0.300 0.275
1.000 1.060 0.400 0.370
2.000 2.200 0.500 0.465
5.000 5.350 0.600 0.560
7.000 7.450 0.700 0.655
10.00 10.60 or 12.5
1)
0.800 0.750
20.00 23.00 0.900 0.845
75.00 83.50 1.000 0.940
1.100 1.035
I
EL
mA I
probe
mA I mA
Limit Value
Maximum
Display Value
Limit Value
Maximum
Display Value
Limit Value
Maximum
Display Value
1.00 0.85 0.100 0.085 0.25 0.12
3.50 3.23 0.250 0.227 0.50 0.35
7.00 6.55 0.500 0.465 1.00 0.80
10.00 9.40 1.000 0.940 2.00 1.70
15.00 14.15 2.000 1.890 3.50 3.05
20.00 18.90 3.500 3.315 5.00 4.40
7.00 6.20
10.00 8.90
15.00 13.40
20.00 17.90
25.00 22.40
65


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUTEST SIII at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUTEST SIII in the language / languages: English as an attachment in your email.

The manual is 0,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info