Automatic Test – Administration
U/M Unit of measure for limit value, setpoint or toler-
ance
Ð Test condition in preparation
Ð MinLimitValue See following table for sample entry
Ð MaxLimitValue See following table for sample entry
Ð Setpoint Ta rg et valu e
Ð Tolerance in preparation
Examples
Note
Select a limit value which either complies with or is stricter
than the legally binding regulations.
Parameter Index Card
The parameters for the measuring function must be entered here. Various
parameters are displayed, depending upon the measuring function. Refer
to section 5, “Manually Triggered Measurements – Individual Measure-
ments” for further information.
Switch off Display / hide message
Switch on Display / hide message
Meas. type e.g. safety class type,
Type of Connection: test socket or permanent
connection
After meas.
1)
Automatic / force test socket off
Polarity L-N or N-L
Up / Ip Test voltage in V / test current in A
fp Test frequency in Hz for I EL
Zero point
2)
Reference point (offset) for R-PE measurement
Duration Test duration in seconds
1)
If these messages are deactivated, attention is drawn to the fact (during
deactivation) that the device may not be switched on via the relays as of
a nominal current of greater than 10 A.
2)
Automatic: It’s determined whether or not line voltage can remain
switched on after measurement. If so, line voltage remains switched on.
This feature can also be deactivated: “force test socket off”. In addition
to this, each measurement which requires line voltage can then be con-
figured such that the inspector is prompted by means of messages to
switch the device on and off after voltage is applied to the test socket,
and before voltage is disconnected from the test socket. This protects
the relay included in the test system.
2)
An offset can be entered for the R-PE measurement in the automatic
test, so that 0 ? is displayed for a selected reference point which is con-
nected to the protective conductor. This is necessary in order to compen-
sate for any offset which is caused by the use of an adapter (e.g. the
AT13/32), or the offset caused by installation in the event of a perma-
nently connected DUT.
Note
Entry of a new measurement can be exited at any time by
pressing the ESC key. Entries are ignored.
Ð Add the new measurement to the catalog by tapping Save .
Test Step Abbr. MinLimit-
Value
Setpoint MaxLimit-
Value
Cross
Reference
Protective conductor meas. R PE 0 0.3 page 45
Insulation resistance meas. R ISO 1 M 10,000 M —page 48
Leakage current I PE 0 mA 3.5 mA page 51
Equivalent leakage current I EL 0 mA 3.5 mA page 54
Contact current I B 0 mA 0.5 mA page 58
Patient leakage current I PL page 61
Eq. patient leakage current I EPL page 63
98


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info