Initial Start-Up – Setup
4 Initial Start-Up
4.1 Connection to the Mains (90 to 264 V, 45 to 65 Hz)
Ð Connect the test system to the mains with the mains plug as
shown in the figure. The mains plug symbol appears in the
header.
Charging the Battery Pack
Ð Initial start-up: Make sure that the battery pack is properly in-
serted and connected (see section 4.2 regarding installation of
the battery pack).
If a battery pack has been installed, it is recharged automatically
as soon as a specified threshold value is fallen short of.
The charging process is indicated by the alternating charge level
at the battery symbol. After this display appears, the test system
should remain connected to the mains for a charging time of at
least 3 hours. Initial charging time: at least 5 hours.
The duration of the charging process is extended correspondingly
in the event of mixed operation, i.e. simultaneous charging and
measuring. Practically no charging at all takes place in the event
of continuous changing from one mains connection to the next
with a dwell time of less than 4 minutes per measuring point. We
recommend charging the battery pack overnight without interrup-
tion. This applies in particular to initial start-up.
You can start fast charging any time with the key (see “Setup
Menu” in section 4.10.8). Even after the batteries have been fully
charged, the device need not be disconnected from mains power
(overcharge protection).
Note
If no battery pack has been installed, the charging symbol
appears at the display after start-up and login.
4.1.1 Automatic Recognition of Mains Connection Errors
The test system automatically recognizes mains connection errors
if the conditions in the following table have been fulfilled. The user
is informed of the type of error, and all measuring functions are
disabled in the event of danger.
Each time the system is reconnected to the mains, the following
messages appear after the first test has been started:
Ð Simultaneously touch the contact surface and tap the “OK”
button.
Ð PE connection test:
If PE is detected, measurement is continued.
If no PE is detected, the following message appears:
“Connect PE”, for which three options exist:
Cancel: Abort the test.
OK: Repeated attempt to detect PE.
Testing is not continued until PE has been detected.
No: Testing is continued regardless of whether or not
PE is detected.
Attention!
!
In the event of mains connection errors in accordance with
either of the first two cases in the following table, immedi-
ately disconnect the test system from the mains and arrange
to have the error eliminated!
Note
Voltage at the electrical system’s protective conductor (PE)
may result in distorted measurement values during testing
for the absence of voltage, or during leakage voltage mea-
surements.
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECUSTAR FM at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECUSTAR FM in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info