GMC-I Messtechnik GmbH 31
Procedure 2. DUT on test socket (new! as from firmware version 7.24)
It is imperative that you observe the specified sequence of test steps:
No device or component may initially be connected to jacks 1 through 3! (Jacks 2
and 3 are short-circuited during all measurements at the test socket;
Exception: as soon as the express command to connect the measurement cable
appears in the display, short-circuit is cancelled, see below)
õ Remove all cables that may be plugged into jacks 1 through 3.
õ Connect the DUT to the test socket.
õ Select switch position VΩ
and then measurement U
AC/DC
.
õ Switch the DUT on (a short-circuit test is conducted).
õ Put the DUT into service by switching line voltage to the test socket
with key (14).
õ Please do not fail to observe the following:
Do not connect the measurement cables
until the following command appears in the display:
„Connect the measurement cable
to jacks 2 and 3 for voltage measurement“.
õ Scan the measuring point with the test probes.
õ Read the measured values.
õ Remove the test probes from the measuring point and pull off the
measurement cables from jacks 2 and 3.
õ Return to the Multimeter menu with ENTER.
Measurement of Safety Extra Low Voltage (see Procedure No. 2, see above)
Line voltage can be switched to the DUT via the test socket with key
14, e.g. in order to measure protective extra low voltage at the DUT output.
Attention!
!
The voltage measured at the DUT output must be a protective
extra low voltage electrically isolated from the mains, otherwise
an overcurrent protection device in the building installation may
trip.
12.3 Resistance R
Resistance of up to 150 kΩ can be measured between the 1 and 2 con-
nector jacks.
Select the R measurement and start.
31


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info