11.3 Patient Leakage Current I
PL
Current which flows from an ap-
plication part via the patient to
earth. This current may be caused
by an unintentional interference
voltage at the patient, and may
flow via the patient and an insu-
lated, floating type F application
part to earth. Useful patient cur-
rent is excluded in both cases.
AC and DC components are measured.
Note
+
Patient leakage current must also be measured if application
parts are available.
The displayed test voltage must be documented.
11.4 Patient Auxiliary Current I
PA
(Feature KA01)
Current which flows within the
patient between the electrodes of
the application part. Use for in-
tended purpose is assumed. Fur-
thermore, the current should not
cause any physiological effects.
For example, this is the case for
input current from amplifiers, or
current used for impedance
plethysmography.
AC and DC components are measured.
11.5 Residual Current I
RC
Sum of instantaneous current values which flow via the L and N conduc-
tors at the device mains connection (also known as differential current).
Residual current is practically identical to fault current in the event of an er-
ror. Fault current: Current which is caused by an insulation defect, and
which flows via the defective point.
11.6 Device Leakage Current I
LC
per IEC 62353 (VDE 0751-1)
Device leakage current is the
sum of all leakage currents from
the housing, all accessible
conductive parts and all applica-
tion parts to PE. Measurement
must be performed for both
mains polarities and the largest
value is documented.
Note
+
The highest device leakage current value and line voltage must
be documented.
Attention!
!
The protective conductor is interrupted during this
measurement.
2
3
(N) L
(L) N
1 Test socket w/o
protective conduc-
tor contact
2 Application part
3 Device housing
4 Measuring circuit
μΑ
4
IEC 62353/
I
LC
1
VDE 0751
27


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info