GMC-I Messtechnik GmbH 25
õ Select the desired nominal voltage
from the display with the key.
The instrument multiplies this nomi-
nal voltage by a factor of 1.5, which
represents the actual test voltage
(see chapter 10.3).
Note: For SC I, the test voltage can
be set to maximum 1.5 kV.
õ Test voltage is applied to the test
socket, and in turn the device under
test, for as long as the
key is
held depressed. Activation of high-
voltage is indicated with an acoustic
signal.
Currently measured minimum output voltage U
HV AC
(measured value
divided by 1.5), test voltage U
DC
, and remaining test time are displayed.
õ Safety class II: Contact all exposed, conductive parts, simultaneously
if possible, in order to avoid unnecessarily long test durations or
repeats. Avoid contacting individual parts sequentially.
õ After the key is released, the equivalent AC voltage is displayed which
occurred during testing. This voltage is indicated as the test result. If
this value is less than the pre-selected nominal voltage, the test is
failed.
õ If you want to repeat the test, press the key. The test sequence is
started over again with selection of nominal voltage.
Attention!
!
If sparkover should occur, the test is immediately interrupted
and the voltage measured at the moment sparkover occurred is
display as U
HV AC
.
If the device is disconnected from the test socket during testing,
even though this is prohibited, the test is immediately inter-
rupted. The following message is displayed: “Caution, device
under test is still charged!”
Testing as Part of a Test Sequence
õ Select the high-voltage menu under the setup function for the respec-
tive test regulation in order to set parameters for the high-voltage test.
õ Enter the desired AC nominal voltages for safety classes I and II. The
instrument multiplies the respective AC nominal voltage by a factor of
1.5, which represents the actual DC test voltage (see chapter 10.3).
Test voltage is determined based upon the selected, or the automati-
cally recognized safety class.
õ Save the setup values to memory.
õ Start the high-voltage test by pressing the key, if “manual
sequence” has been selected in the setup menu.
õ Safety class II only: Contact the device under test with the probe.
Activation of high-voltage is indicated with an acoustic signal.
The selected nominal voltage of 3.5 kV results in a DC output voltage of
maximum 5.25 kV.
The high-voltage test is ended auto-
matically after the test duration has
elapsed.
U
HV DC
: Equivalent test voltage in DC
U
HV AC
: Measured DC value divided
by 1.5
If the value U
HV AC
is less than pre-se-
lected nominal voltage U
AC
, the test is
failed.
Note
+
In the event of sparkover, the voltage measured at the moment
sparkover occurred is displayed in the test results as minimum
value U
HV AC
, along with the reason the test was failed.
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ST HV at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ST HV in the language / languages: English as an attachment in your email.

The manual is 0,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info