Control Functions
4.4 Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with
the DATA function (auto-hold). This is useful, for example, when
contacting the measuring points with the test probes requires
your full attention. After the measuring signal has been applied
and the measured value has settled in in accordance with the
condition” listed in the table below, the measured value is frozen
at the digital display and an acoustic signal is generated. The test
probes can now be removed from the measuring points, and the
measured value can be read from the digital display. If the
measuring signal falls below the value specified in the table, the
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by
less than 100 digits, the acoustic signal is generated twice. If
deviation is greater than 100 digits, only a brief acoustic signal is
generated.
Note
The DATA function has no effect on the analog display, at
which the current measured value continues to appear.
However, when the digital display if “frozen”, the decimal
point is fixed as well (fixed measuring range, symbol:
MAN).
The selected measuring range should not be manually
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the
DATA/MIN/MAX key (for approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
1)
Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is
saved as a reference value. For subsequent data hold, two acoustic
signals are only generated if the currently frozen value deviates from the
first saved value by less than 100 digits.
Key: MV = measured value, R = measuring range
DATA
Function
Press
DATA /
Min/Max
Condition Response from Instrument
Measuring
Function
Measuring
Signal
Display
Acoustic
MV
Digital
DATA
Activate Brief blinks Once
Save
(stabilized
measured
value)
V, A, F, Hz,
MHz, %
> 10%
of R
Is dis-
played
Static
Once
Twice
2)
Ω
0L
Reactivate
1)
V, A, F, Hz,
MHz, %
< 10%
of R
Stored
MV
Blinks
Ω
= 0L
Change to
Min/Max
Brief See table in chapter 4.4.1
Exit Long
Is cleared Is cleared
Twice
V, A, Hz
t [s]
100%
10%
Activated
Reactivated
Save
Save
10,000 digits
of measuring range
F, MHz, %
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE HIT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE HIT in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt SECULIFE HIT

Gossen Metrawatt SECULIFE HIT User Manual - Dutch - 70 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info