18 GMC-I Messtechnik GmbH
Control Functions
4 Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
The rotary switch is linked to the automatic socket blocking
mechanism, which only allows access to two connector jacks for
each function. Be certain to remove the appropriate plug from its
respective jack before switching to and from the “Afunctions.
The socket blocking mechanism prevents the user from inadvert-
ently turning the selector switch to impermissible functions after
the measurement cables have been plugged in to the instrument.
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, diode and
continuity testing, and the MHz measuring function. Auto-ranging
is active as soon as the instrument is switched on. The instrument
automatically selects the measuring range which allows for
highest possible resolution of the applied quantity. When the
instrument is switched to frequency measurement, the previously
selected voltage measuring range remains active.
The Auto-Ranging Function
The multimeter is switched automatically to the next higher range
at ±(11999 d + 1 d 01200 d), and to the next lower range at
±(01100 d - 1 d 10990 d).
Exception, capacitance measurement:
The multimeter is switched automatically to the next higher range
at ±(1199 d + 1 d 0120 d), and to the next lower range at
±(0110 d - 1 d 1099 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
or
scroll key.
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
* Via manual measuring range selection only
Function Display
MAN / AUTO
Manual mode active:
utilized measuring range is fixed
MAN
or
Range switching sequence for:
V: 100 mV*
1 V
10 V
100 V
1000 V
Hz: 100 Hz
1 kHz
10 kHz
100 kHz
Ω: 100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
40 MΩ
A: METRAHITX-
TRAOUTDOOR:
100
μ
A
1 mA
10 mA
100 mA
1 A
10 A (16 A)
METRAHITT
ECH:
10 mA
100 mA
1 A
10 A (16 A)
METRAHITP
RO: 1A
10 A (16 A)
A:METRAHITB
ASE and METRAHITTECH:
see chapter 5.7.3 ff
F: METRAHITX-
TRAOUTDOOR and METRAHITTECH:
10 nF
100 nF
1
μ
F
10
μ
F
100
μ
F
1000
μ
F
MAN
MAN / AUTO
Return to automatic measuring range selection
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE HIT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE HIT in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt SECULIFE HIT

Gossen Metrawatt SECULIFE HIT User Manual - Dutch - 70 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info