590784
124
Zoom out
Zoom in
Previous page
1/130
Next page
124 GMC-I Messtechnik GmbH
SECULIFEESXTRA
The RF Measure screen shows mA, but I
need to measure watts. Can the
SECULIFE ES
XTRA measure watts?
Yes, press the mA RMS text to select the measurement to be shown.
You can also press the Display Parameters button to select how
many measurements are shown on the screen at a time.
Why do I have two blue REM cables?
The cable with the clear pin is used to enable the dual pad function
(REM) on the DUT. This cable is used for REM test mode only. For
RF Power measurements the SECULIFE ES
XTRA shorts the two
dispersive inputs internally. The blue cable without the pin is used
because it disables the REM function on the DUT.
I am testing a ForceTriad. When I key the
generator, I get a REM alarm on the
ForceTriad.
You must configure the DUT for Demo mode to bypass the REM
alarm. Use the Tool icon on the right most display and select Demo
Mode from the menu on the left most display.
The SECULIFE ES
XTRA shows a mV
RMS measurement. Is this the load
voltage?
No, this is the output voltage of the internal Current to Voltage RF
transformer. This was included for compatibility with the SECULIFE
ESPRO analyzer. To get load voltage, multiply the RMS current by
the load resistance.
Question Answer
124


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ES XTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ES XTRA in the language / languages: English as an attachment in your email.

The manual is 7,06 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info