590784
123
Zoom out
Zoom in
Previous page
1/130
Next page
GMC-I Messtechnik GmbH 123
SECULIFEESXTRA
FREQUENTLY ASKED QUESTIONS (FAQ)
Question Answer
What are the new features in the 2.0.3.15
software?
There are many new features, see the User's manual for details.
Added secure autosequence for ForceTriad PM procedure
Option to select details or summary data when saving Autose-
quence results to PDF files
RS-232 terminal window shows mA measurement and can set
load
New secure autosequence, once saved as secure sequence, the
settings can't be modified by user
Copy Step function added to Autosequence configuration
Changed colors to help identify RF Measure and RF Leakage
screens
Added ground referenced measurement mode for RF Measure
screen
Added network drive access to file open/save
Added sort by filename or date to file open/save
I am trying to take a measurement, why do
the readings frequently bounce to zero?
Chances are you are trying to read a pulsed waveform. Try setting
the measurement mode to 'Pulsed' instead of 'Continuous'.
How do I measure a bipolar output when
there is only a REM and Active cable
provided?
Initial units were only shipped with two active leads, Green and
Yellow. Use the Yellow active cable to connect from the Active port
on the SECULIFE ES
XTRA to one bipolar output on the DUT. Use
the Green cable to connect from Dispersive 1 on the SECULIFE
ES
XTRA to the other bipolar output on the DUT.
Starting with Rev G units, a Blue wire was added to the accessory kit.
The Yellow and Blue wires can be used to measure bipolar outputs.
Why aren't there more pre-programmed
autosequences provided with the
SECULIFE ES
XTRA?
We have been restricted by the ESU manufacturers from including
autosequences with the SECULIFE ES
XTRA. However, the
Quickstart guide can help with configuring autosequences.
I just want to use the SECULIFE ES
XTRA
to measure current and don't need all of
the other features. Is there an easy way to
do this?
Yes, you can use profiles to configure the SECULIFE ES
XTRA to boot
to any screen. Select System Tools from the main screen, then
System Setup. The Startup Screen can then be selected and saved
to a profile.
How do I enter text for the Autosequence
title or instructions to user?
Click in the text box to be edited. To enter text, you will need an
external USB or PS/2 keyboard. You can also enter the text from
your PC using the Remote mode.
What is the tissue test, how is it used?
Some generators have the ability to seal tissue and detect when the
seal is made. These generators use a changing load impedance to
detect the seal. The tissue response test uses a light bulb as a load
to simulate changing tissue impedance. This mode will monitor the
current and report the value right before the output is turned off.
What is pulsed input mode for? Isn't Coag
a pulsed waveform?
Pulsed mode is for measuring outputs that have a relatively long duty
cycle such as 20mS off, 500mS on. Typically, the generator will
have a special button to enable the pulsed output mode.
123


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ES XTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ES XTRA in the language / languages: English as an attachment in your email.

The manual is 7,06 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.