SECULIFE ESPRO
Typical Setup Screen
The following is a breakdown of the parameters available in the configuration of the unit and their
available options:
Factory Setup Configuration
Parameter Description Range
Input Range
Used to select the input range that is to be
calibrated.
100 mV
1000 mV
Donut Attenuation
Selects the Attenuation in Volts : Amp for the RF
Current Transformer
being used. Each setting
has an independent gain setting.
0.1 : 1
1 : 1
Volts : Amps
Counts Zero Offset
Used to adjust the A/D converter output to zero
counts when zero input is applied. This should
be adjusted to get the smallest mVRMS reading
when the input is zero.
1000 Counts
Input Gain
Used to calibrate the mV reading for when the
0.1 : 1 RF Current Transformer
(Donut)
attenuation is selected.
0 to 5000
mV RMS Reading
Shows the latest measurement by the system.
This is for display only to help in calibrating the
system.
0.0 to 700.0 mV
Input Zero
Zeros the input circuitry based on donut being
used. Each donut can have a slightly different
zero offset. This function will eliminate the offset
from the readings. Press the SETUP key while
this parameter is selected to perform the auto –
zero function. This will also remove any offsets
created by the User in the System Setup screen.
Press Setup
28 GMC-I Messtechnik GmbH
28


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt SECULIFE ES PRO at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt SECULIFE ES PRO in the language / languages: English as an attachment in your email.

The manual is 0,89 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info