590777
87
Zoom out
Zoom in
Previous page
1/96
Next page
GMC-I Messtechnik GmbH 87
20 Maintenance
20.1 Firmware Revision and Calibration Information
See section 4.6.
20.2 Rechargeable Battery Operation, and Charging
Check to make sure that no leakage has occurred at the
rechargeable batteries at short, regular intervals, or after the
instrument has been in storage for a lengthy period of time.
Note
Prior to lengthy periods of rest (e. g. holiday), we recom-
mend removing the rechargeable batteries. This helps to
prevent excessive depletion or leakage of batteries,
which, under unfavourable circumstances, may cause
damage to the instrument.
If battery voltage has fallen below the allowable
lower limit, the pictograph shown at the right
appears. “Low Batt!!!” is also displayed along with a battery icon.
The instrument does not function if the batteries have been
depleted excessively, and no display appears.
Attention!
!
Use only the charger Z502R to charge the Kompakt Akku-
Pack Master (Z502H) which has already been inserted into
the test instrument.
Make sure that the following conditions have been fulfilled be-
fore connecting the charger to the charging socket:
– Kompakt Akku-Pack Master (Z502H) has been
installed, no commercially available battery packs, no
individual rechargeable batteries, no standard batteries
– The test instrument has been disconnected from the
measuring circuit at all poles
– The instrument must remain off during charging.
If the batteries or the battery pack (Z502H) have not been used or
recharged for a lengthy period of time (> one month), thus result-
ing in excessive depletion:
Observe the charging sequence (indicated by LEDs at the char-
ger) and initiate a second charging sequence if necessary (dis-
connect the charger from the mains and from the test instrument
to this end, and then reconnect it). Please note that the system
clock stops in this case and must be set to the correct time after
the instrument has been restarted.
20.2.1 Charging Procedure with Charger for Z502R
Insert the correct mains plug for your country into the charger.
Attention!
!
Make sure that Kompakt Akku Pack Master (Z502H) has
been inserted, no battery holder.
For charging in the tester, only use Kompakt Akku Pack Master (Z502H),
which is either included in the standard equipment or available as an
accessory, with heat-sealed battery cells.
Connect the charger to the test instrument with the jack plug,
and then to the 230 V mains with the interchangeable plug.
(The charger is suitable for mains operation only!)
Attention!
!
Do not switch the test instrument on during charging.
Monitoring of the charging process by the microproces-
sor might otherwise be disturbed, in which case the
charging times specified in the technical data can no lon-
ger be assured.
Please refer to the operating instructions included with the
charger regarding the meanings of LED displays during the
charging process.
Do not disconnect the charger from the test instrument until
the green LED (charged/ready) lights up.
20.3 Fuses
If a fuse has blown due to overload, a corresponding message
error appears at the display panel. The instrument’s voltage mea-
suring ranges are nevertheless still functional.
Replacing the Fuse
Attention!
!
Disconnect the device from the measuring circuit at all
poles before opening the fuse compartment lid!
Loosen the slotted screws at the fuse compartment lid next to
the mains power cable with a screwdriver. The fuses are now
accessible.
Replacement fuses can be accessed after opening the battery
compartment lid.
Attention!
!
Severe damage to the instrument may occur if incorrect
fuses are used.
Only original fuses from GMC-I Messtechnik GmbH may
be used (order no. 3-578-285-01 / SIBA 7012540.3.15
SI-EINSATZ FF 3.15/500 6.3X32).
Only original fuses assure required protection by means
of suitable blowing characteristics. Short-circuiting of
fuse terminals or the repair of fuses is prohibited, and is
life endangering!
The instrument may be damaged if fuses with incorrect
ampere ratings, breaking capacities or blowing charac-
teristics are used!
Remove the defective fuse and insert a new one.
Insert the fuse compartment lid after the fuse has been re-
placed and secure it by turning clockwise.
20.4 Housing
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. In par-
ticular for the protective rubber surfaces, we recommend a moist,
lint-free microfiber cloth. Avoid the use of cleansers, abrasives or
solvents.
Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German electrical and
electronic device law). This device is subject to the RoHS direc-
tive. Furthermore, we make reference to the fact that the current
status in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search term WEEE.
In accordance with WEEE 2012/19EU and ElektroG, we
identify our electrical and electronic devices with the sym-
bol in accordance with DIN EN 50419 which is shown at
the right. Devices identified with this symbol may not be disposed
of with the trash. Please contact our service department regard-
ing the return of old devices (see address in section 22).
If the (rechargeable) batteries used in your instrument are depleted,
they must be disposed of properly in accordance with valid
national regulations.
Batteries may contain pollutants and heavy metals such as lead
(Pb), cadmium (Cd) and mercury (Hg).
The symbol to the right indicates that batteries must not
be disposed of with the trash, and must be brought to a
designated collection point.
BAT
Pb Cd Hg
87


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt PROFITEST MXTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt PROFITEST MXTRA in the language / languages: English as an attachment in your email.

The manual is 2,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt PROFITEST MXTRA

Gossen Metrawatt PROFITEST MXTRA User Manual - Dutch - 92 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info