GMC-I Messtechnik GmbH 55
14.4 Leakage Current Measurement
with PRO-AB Leakage Current Adapter as Accessory
– I
L
Function (PROFITEST MXTRA & SECULIFE IP only)
Applications
Measurement of contact voltage in accordance with DIN VDE
0107, part 10, as well as continuous leakage and patient auxiliary
current per IEC 62353 (VDE 0750, part 1) / IEC 601-1 / EN 60
601-1:2006 (Medical electrical equipment – General requirements
for basic safety), is possible using the PRO-AB PRO-AB leakage
current measuring adapter as an accessory with the
PROFITEST MXTRA test instrument.
As specified in the standards listed above, current values of up to
10 mA may be measured with this measuring adapter. In order to
be able to fully cover this measuring range using the measure-
ment input provided on the test instrument (2-pole current clamp
input), the measuring instrument is equipped with range switching
between transformation ratios of 10:1 and 1:1. In the 10:1 range,
voltage dividing takes place at the same ratio.
Connection and Test Set-Up
In order to perform the leakage current measurement, the
adapter’s measurement outputs must be plugged into the mea-
surement inputs at the left-hand side of the PROFITEST MXTRA (2-
pole current clamp input and probe input).
Either of the leakage current measuring adapter’s inputs is con-
nected to reference earth (e.g. safe earth electrode / equipotential
bonding) via a measurement cable. The metallic housing (accessi-
ble part) of the device under test is contacted with a test probe or
alligator clip which is connected to the other input by means of a
second measurement cable.
Testing the PRO-AB Adapter
The adapter should be tested before use and at regular intervals
(see adapter operating instructions).
Measuring Sequence
Refer to the operating instructions for the PRO-AB leakage cur-
rent measuring adapter regarding performance of the measure-
ment.
Attention!
!
The test plug should be located in the storage slot during
leakage current measurement. Under no circumstances
may the test plug be connected with any system compo-
nents, including PE / ground potential (measured values
might otherwise be distorted).
The measurement can be started or stopped by pressing the
“START” key. Leakage current measurement is a long-term mea-
surement, i.e. it continues until it is stopped by the user. The
momentary measured value is display continuously during mea-
surement.
Note
The self-test must be deactivated in the menu (set
TEST ON/OFF” function key to “OFF”) in order to perform a
measurement.
Always start with the large measuring range (10:1), unless there’s
no doubt that small measured values can be expected, in which
case the small measuring range can be used (1:1). The measuring
range must be selected at the measuring adapter, as well as in
the menu using the corresponding function key (RANGE). It must
be assured that the range settings at the adapter and at the test
instrument are always identical, in order to prevent any distortion
of measurement results.
Depending on the magnitude of the measured values, the range
setting can, or must (in the case of overranging), be manually cor-
rected at the measuring adapter and the test instrument.
Individual limit values can be adjusted after pressing the “Limits”
function key. Exceeded limit values are indicated by the red limit
value LED at the test instrument.
55


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt PROFITEST MXTRA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt PROFITEST MXTRA in the language / languages: English as an attachment in your email.

The manual is 2,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt PROFITEST MXTRA

Gossen Metrawatt PROFITEST MXTRA User Manual - Dutch - 92 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info