GMC-I Messtechnik GmbH 27
Display Values in Consideration of Measuring Uncertainty
Table for determining minimum display values for insulation resis-
tance in consideration of the instrument’s measuring uncertainty.
Table for determining maximum display values for low-value resis-
tance in consideration of the instrument’s measuring uncertainty.
Voltage at Device Under Test During Insulation Resistance
Measurement
Measuring voltage U
x
at the device under test depending upon its
resistance R
x
at nominal voltage U
N
= 100, 250, 500 and 1000 V:
Limit Value
Minimum
Display Value
Limit Value
Minimum
Display Value
020 kΩ 025 kΩ
100 kΩ 111 kΩ 100 MΩ 111 MΩ
200 kΩ 219 kΩ 200 MΩ 219 MΩ
500 kΩ 541 kΩ 500 MΩ 541 MΩ
0.20 MΩ 0.25 MΩ
0.50 MΩ 0.57 MΩ
1.00 MΩ 1.11 MΩ 1.00 GΩ 1.11 GΩ
2.00 MΩ 2.19 MΩ 2.00 GΩ 2.19 GΩ
5.00 MΩ 5.41 MΩ 5.00 GΩ 5.41 GΩ
10.0 MΩ 11.1 MΩ 10.0 GΩ 11.1GΩ
20.0 MΩ 21.9 MΩ 20.0 GΩ 22.6 GΩ
50.0 MΩ 54.1 MΩ 50.0 GΩ 55.9 GΩ
Limit Value
Maximum
Display Value
Limit Value
Maximum
Display Value
0.15 Ω 0.11 Ω
0.20
Ω 0.16 Ω 5.00 Ω 4.72 Ω
0.50 Ω 0.44 Ω 10.0 Ω 9.47 Ω
1.00 Ω 0.92 Ω 20.0 Ω 17.7 Ω
2.00 Ω 1.87 Ω 50.0 Ω 44.7 Ω
100
U
x
/V
R
x
/kΩ
U
N
= 100 V
0
20
40
60
80
100
250
U
x
/V
R
x
/kΩ
U
N
= 250V
0
50
100
150
200
250
27


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO TECH at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO TECH in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO TECH

Gossen Metrawatt METRISO TECH User Manual - German - 6 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info