54 GMC-I Messtechnik GmbH
7 Characteristic Values
Standards DIN EN 61557-1:2007
DIN EN 61557-2:2008
VDE regulations VDE 0413 Part 1:2007
VDE 0413 Part 2:2008
Insulation Resistance
Test duration: automatic (until measured value stabilizes),
manual (1 ... 120 s) or continuous measurement (lock function)
Polarization Index (PI), Absorption Ratio (DAR)
PI and DAR are calculated values. Insulation measurement specifications apply.
Insulation Test Voltage
Variable test voltage selectable in 50 V steps
Short-circuit current less than 2 mA with test voltage to 1.00 kV
Voltage Measurement
Frequency Measurement
Measured quantity voltage: 50 V ... 1 kV
Breakdown Voltage
Display Range
[Ω]
Measuring
Range [Ω]
Test Voltage Intrinsic Uncer-
tainity
Measuring Uncer-
tainty
0.00 M ... 50.0 G 0.60 M ...10.0 G 100 V ...250 V ±(10% rdg. + 5 d) ± (15% rdg. + 7 d)
>10.0 G...50.0 G ±(25% rdg. + 5 d) ± (50% rdg. + 20 d)
0.00 M ... 250 G 0.40 M...50.0 G >250 V ... 1.00 kV ±(7% rdg. + 5 d) ± (10% rdg. + 7 d)
>50.0 G...250 G ±(25% rdg. + 5 d) ±(50% rdg. + 20 d)
0.00 M ... 999 G 0.40 M...200 G >1.00 kV ... 5.00 kV ±(7% rdg. + 5 d) ±(10% rdg. + 7 d)
>200 G...999 G ±(25% rdg. + 5 d) ±(50% rdg. + 20 d)
t1 t2 Limit
PI
00:00 ... 01:00 ... 99:50 min 00:00 ... 10:00 ... 99:50 min 0.10 ... 4.00 ... 9.80 min
DAR
00:00 ... 00:30 ... 99:50 min 00:00 ... 01:00 ... 99:50 min 0.10 ... 1.60 ... 9.80 min
Nominal Values,
Test Voltage
Variable
Test Voltage
Nominal Current Intrinsic Uncertainty
100 V, 250 V, 500 V, 1.00 kV
1.0 mA 0 ... +25% rdg.
1.50 kV, 2.00 kV, 2.50 kV 0.4 mA ± 5% rdg.
5.00 kV 0.1 mA ± 3.5% rdg.
100 V...1.00 kV 1.0 mA ± 15% rdg.
> 1.00 kV...2.50 kV 0.4 mA ± 5% rdg.
> 2.50 kV...5.00 kV 0.1 mA ± 3.5% rdg.
Measuring Range Frequency
in Hz
Impedance Intrinsic
Uncertainty
Measuring
Uncertainty
dc test voltage
50 V ... 5.00 kV
±(2.5% rdg. + 5 d) ±(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc 15 ... 500 1 MΩ ±(2.5% rdg. + 2 d) ± (5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc >500 ... 1 k 1 MΩ ±(10% rdg. + 2 d)
±(12.5 % rdg. + 5 d)
Measuring Range Impedance Intrinsic
Uncertainty
Measuring
Uncertainty
15.0 Hz 1.00 kHz 1 MΩ±(0.5% rdg. + 2 d) ±(1% rdg. + 2 d)
Parameter Setting Range Intrinsic Uncertainty Measuring
Uncertainty
Voltage range 100 5,000 V ± (10% rdg. + 8 d)
±
(
15% rdg. + 10 d)
Rise time 5 300 s
Measuring time 1 120 s /
auto / continuous
measurement
——
54


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME plus in the language / languages: English as an attachment in your email.

The manual is 1,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME plus

Gossen Metrawatt METRISO PRIME plus User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info