590655
66
Zoom out
Zoom in
Previous page
1/68
Next page
METRISO PRIME+
Digital High-Voltage Insulation Test Instrument
Operating Instructions
Test instrument may
only be operated under
the supervision of a
qualified electrician!
3-349-822-03
2/3.15
2 GMC-I Messtechnik GmbH
12345
6
7
8
9 10
13
12
12
15
11
18
16
19
17
11
14
GMC-I Messtechnik GmbH 3
1 Function selector switch
Switch positions:
SETUP: General device settings
Print out measured values and reports
Select data memory function
OFF: Switches the test instrument off *
TEST: Select measuring and test parameters,
and execute measurements and tests
2 : Select main menus and submenus, and adjust values
3 : Select main menus and submenus, and adjust values
4 Socket connector for RS 232 interface
5LCD panel
6 Signal lamp for test passed (green) / test failed (red)
7 MENU: Key for selecting and querying menus for setting
parameters, for returning to the main menu and for
stopping the measuring process
8 i/STORE: Before performing a measurement:
Key for querying context sensitive help
After performing a measurement:
Key for saving measured values and parameters to memory
9 START: Key for starting the selected test
10 Signal lamp: “testing ...” (yellow)
11 Pushbuttons (left and right) for releasing the handle
12 Pushbuttons (left and right) for releasing the cover
13 Cover
14 SECUTEST PSI printer module
15 Handle and bar function as a tilt stand
* During battery operation: battery is disconnected
During mains operation: transformer is still connected to the mains
16 Inlet plug for supply power with fuse compartment
17 Connector jack for guard cable (feature G1)
18 Connection for automotive charging adapter J1
19 Device base
Outside: With uptake for measurement cable, test probes and alligator clips
Inside: Uptake for guard cable with test probe (feature G1),
battery compartment
Attention!
!
Connect 5 m long Leadex 5000 accessory extension
cable (feature H1) to + measurement cable only!
Note
Maximum rated voltage to earth is 1000 V~, even if perma-
nently connected measurement cables for 2 kV~ CAT II
are permissible.
4 GMC-I Messtechnik GmbH
Retrofitting with a high-voltage module (feature B1 or B2) can only
be performed by GMC-I Service GmbH.
The high-voltage module is
then controlled by MERISO PRIME+.
1 Inlet plug for high-voltage module supply power with fuse insert
2 Key switch for prevention of unauthorized start-up
3 Connection for external signal lamps (see chapter 3.2.3 on page 18)
4 High-voltage module ready for operation when signal lamp is lit up green
5 High-voltage module ready to switch on when signal lamp is lit up red
Caution, danger!
6 Permanently connected measurement cables
7Trigger (switch)
As opposed to the high-voltage pistol without switch, the high-voltage
switching pistol is identified with a red clamping ring on the connector cable
underneath the handle.
8 Recessed safety test probe
0
I
!
5
4
1
2
3
PROFITEST 204 HP/HV High-Voltage Module,
Connection Panel
6
8
7
High-Voltage Switching Pistol
GMC-I Messtechnik GmbH 5
PC Program WinProfi for communication with MERISO PRIME+
The free PC starter software WinProfi is used for communication
with your MERISO PRIME+ test instrument. WinProfi* is available
on our homepage with the following content and functions:
up-to-date test instrument software
– for loading a different language
for the user interface
– for loading an updated software version
Transmission of measurement data from the test instrument to
the PC
preparation and modification of templates for test reports at the
PC and transmission of the templates to the test instrument
preparation, printout and filing of test reports at the PC
The following interface cables and/or converters are required for
communication between test instrument and PC:
Z3241 RS232 (tester*) – RS232 (PC)
RS232-USB Converter (Z501L) RS232 (tester*) – USB (PC)
* not possible via the RS232 terminal of a connected SECUTEST PSI
Up-to-date PC software (free of charge starter or demo software
for data management, report and list generation) is available from
our homepage for download.
* WinProfi is currently applicable up to Windows 7 (32 Bit)
Contents Page Contents Page
6 GMC-I Messtechnik GmbH
1 Applications ..................................................................................8
1.1 List of Available Features ...............................................................9
2 Safety Features and Precautions ................................................10
2.1 Checklist for High-Voltage Tests ....................................................11
2.2 Significance of Symbols .................................................................12
2.2.1 Symbols on the Device ..............................................................12
2.2.2 Symbols in the Operating Instructions .........................................12
2.2.3 Symbols in the User Interface ....................................................12
2.2.4 Symbols in the User Interface with Feature B1/B2 .......................13
2.3 Special Safety Precautions and Instructions
for Feature B1/B2 .........................................................................13
3 Initial Start-Up ............................................................................15
3.1 Battery Operation, Feature C1 (not with B1/B2) ...............................15
3.1.1 Removing the Device Base and Inserting or Replacing
the Battery Pack .......................................................................15
3.2 Initial Start-Up with Feature B1/B2 .................................................17
3.2.1 Retrofitting Feature B1/B2 .........................................................17
3.2.2 Key Switch (feature B1 or B2) ....................................................17
3.2.3 Indication of Operating States ....................................................18
3.3 Mounting the MERISO PRIME+ with Feature B1/B2 to the Trolley (feature D1).
................................................................................................................ 19
3.4 Mains Connection .........................................................................20
3.4.1 DC Operation (automotive charging adapter in preparation) ..........20
3.5 User Interface ...............................................................................21
3.6 Help Functions ..............................................................................21
3.7 Setup ...........................................................................................22
3.7.1 Adjusting Contrast and LCD Illumination .....................................22
3.7.2 Setting Date and Time ...............................................................23
3.7.3 Beeper Settings .......................................................................24
3.7.4 Executing the Self-Test ..............................................................25
3.7.5 Downloading a Language or a Software Update .........................26
4 Performing Tests ........................................................................29
4.1 Insulation (resistance) Test ............................................................29
4.1.1 Setting Test Parameters ............................................................30
4.1.2 Starting the Test (U
NOM
, U
VAR
) ...................................................31
4.1.3 Starting the Test (U
STEP
) ............................................................31
4.1.4 Data logger function (min log)(as of firmware version AH) ............32
4.1.5 Measurements with the Guard Cable (feature G1) ......................33
4.2 High-Voltage Test with Feature B1/B2 ............................................34
4.2.1 Performance Test (preparation for testing) ..................................34
4.2.2 Setting Test Parameters for High-Voltage Testing .......................36
4.2.3 High-Voltage Testing Sequence .................................................37
4.2.4 Pulse Control Mode ...................................................................39
4.2.5 Ending the High-voltage Test .....................................................39
4.3 Polarization Index Measurement ...................................................40
4.3.1 Setting Test Parameters ............................................................41
4.3.2 Starting the Test .......................................................................41
4.4 Breakdown Voltage ......................................................................42
4.4.1 Setting Test Parameters ............................................................43
4.4.2 Starting the Test .......................................................................43
4.5 Capacitance Measurement and
Determination of Dielectric Discharge (optional)
................................. 44
4.5.1 Setting Test Parameters ............................................................44
4.5.2 Starting the Test .......................................................................45
4.6 Voltage Measurement (protection against residual voltage) ............46
4.6.1 Setting Test Parameters ............................................................47
4.6.2 Starting the Test .......................................................................47
GMC-I Messtechnik GmbH 7
Contents Page Contents Page
5 Processing, Reorganizing and Deleting Data ............................. 48
5.1 Select Device under Test (DUT) .................................................... 48
5.1.1 Entering Descriptions ............................................................... 49
5.1.2 Copying Descriptions ............................................................... 50
5.1.3 Deleting Descriptions ............................................................... 50
5.2 Processing Data (checking measured values) ................................ 50
5.3 Data Reorganization ...................................................................... 51
5.3.1 Deleting Already Printed Data ................................................... 51
5.3.2 Memory Test ........................................................................... 51
5.4 Clearing the Memory ................................................................... 52
6 Printing Test Results ................................................................. 53
7 Characteristic Values ................................................................. 54
8 Data Interfaces ........................................................................... 57
8.1 RS 232 Serial Interface ................................................................. 57
8.1.1 Software Analysis of Measurement Results ................................ 57
8.1.2 Interface Specification and Protocol ........................................... 57
8.1.3 Connector Pin Assignments ....................................................... 57
9 Displays and Error Messages – Causes – Remedies ................ 58
10 Maintenance .............................................................................. 61
10.1 Replacing the Mains Fuse ............................................................. 61
10.2 Testing of Rechargeable Batteries ................................................. 61
10.3 Housing and Test Probes .............................................................. 61
10.4 Measurement Cables .................................................................... 62
10.5 High-Voltage Module Test Cables (Feature B1/B2) .......................... 62
10.6 Replacing Bulbs in the External Signal Lamps (Feature F1) .............. 62
10.7 Software ...................................................................................... 62
10.8 Recalibration ................................................................................ 63
10.9 Device Return and Environmentally Compatible Disposal ................. 63
11 Appendix .................................................................................... 64
11.1 Glossary ..................................................................................... 64
11.2 Minimum Display Values in Consideration of
Measuring Error ........................................................................... 64
11.3
Voltage at Device Under Test During Insulation Resistance Test ...................65
11.4 Index ........................................................................................... 66
12 Repair and Replacement Parts Service,
Calibration Center and Rental Instrument Service .................................67
13 Product Support ......................................................................... 68
8 GMC-I Messtechnik GmbH
1 Applications
The MERISO PRIME+ insulation measuring instrument is in com-
pliance with the following standards and regulations:
Standard DIN EN 61557-1:2007
DIN EN 61557-2:2008
VDE regulation VDE 0413 Part 1:2007
VDE 0413 Part 2:2008
“Equipment for testing, measuring or monitoring protective mea-
sures in systems with line voltages of up to 1000 V AC and
1500 V DC,
part 2, insulation resistance measuring instruments”.
The instrument is suitable for the measurement of insulation resis-
tance at voltage-free devices and equipment with nominal volt-
ages of up to 1000 V. It is also suitable for the measurement of
insulation resistance with values of up to 1 TΩ, and open-circuit
voltages of up to 5000 V.
The instrument is also equipped with a 1000 V measuring range
for direct and alternating voltages. This provides users with the
advantage of checking devices under test for the absence of volt-
age, and discharging capacitive devices under test.
All of the values required for approval reports can be measured
with this instrument.
The SECUTEST PSI module (feature I1) – a printer which can be
inserted into the cover which includes an integrated interface and
keypad – expands the range of possible applications for the
MERISO PRIME+.
All measured data can be archived with the help of measurement
and test reports which can be printed out directly via a PC. This is
a very important feature, especially with regard to product reliabil-
ity.
Notes Regarding SECUTEST PSI Operating Instructions
In combination with the MERISO PRIME+, the SECUTEST PSI
printer module is only used for printing out measured values and
for entering comments with the alphanumeric keypad.
The following chapters from the SECUTEST PSI operating
instructions are relevant as regards operation together with the
MERISO PRIME+:
Chapter 2 Safety Features and Precautions
Chap. 3.1 and 3.2Installing the Batteries and the PSI Module
Chapter 12 Characteristic Values
Chapter 13 Maintenance
Chapter 14 Repair and Replacement Parts Service
The following functions are not active, or only make sense in com-
bination with a SECUTEST test instrument:
Display, print and store reports to memory
Statistics
Operation with a barcode scanner
GMC-I Messtechnik GmbH 9
1.1 List of Available Features
1)
If the high-voltage module is purchased after the basic instrument, it can only be
retrofitted by our service department.
2)
Only for line voltage 207 V 253 V / 49 Hz 61 Hz
3)
The test instrument can be recalibrated by our calibration service at any time.
We recommend a calibration interval of 1 to 2 years.
Specify the designation of the basic M5000 instrument in your order, as
well as any features which deviate from feature 0!
Example of a complete type designation
(= article number, = order code) for a MERISO PRIME+:
Test instrument for German speaking countries with DAkkS
calibration certificate and SECUTEST PSI printer module:
M5000 A01 E1 I1
Features 0 0102040507101543
Country version
(user interface
language,
mains plug type)
A
D
GB in-
ter-
nati-
onal
FRA
F
NLD
NL
ESP
E
ITA
I
CHE
CH
USA
USA
PROFITEST204HP/2.5kV
(not with C1)
B
1)
none
with
PROFITEST204HV/5.4kV
(not with C1)
B
1)
none
with
2)
Set of batteries
(not with B1, B2)
C none
with
DAkkS calibration
certificate
3)
E none
with
“Guard 5000A
measurement cable
G none
with
“LEADEX 5000”
extension cable
H none
with
SECUTEST PSI
printer module
I none
with
10 GMC-I Messtechnik GmbH
2 Safety Features and Precautions
The MERISO PRIME+ test instrument, and the
PROFITEST204HP/2.5kV and PROFITEST204HV/5.4kV high
voltage modules have been manufactured and tested in accor-
dance with the following safety regulations:
Standard DIN EN 61557-1:2007
DIN EN 61557-2:1998-05
IEC 61010-1:2004
EN 61010-1:2002
VDE regulation VDE 0413 Part 1:2007
VDE 0413 Part 2:1998-05
VDE 0411 Part 1:2002
If used for its intended purpose, safety of the user, the instrument
and the device under test is assured.
Read the operating instructions completely and carefully before using the
device, and follow all instructions included therein.
The operating instructions should be made available to all users.
Observe the following safety precautions:
The instrument may only be connected to a 230 V or 120 V
electrical system (depending on the country-specific design)
which is protected with a fuse or circuit breaker with a maxi-
mum nominal current rating of 16 A.
Be prepared for the occurrence of unexpected voltages at de-
vices under test (for example, capacitors may be dangerously
charged).
Make sure that connector, measurement and test cables are
not damaged (i.e. broken insulation, kinks, interruptions etc.).
The instrument may only be used indoors.
Impulse withstand voltage of up to 2.5 kV may occur in
measurement category II systems with 230/400 V.
If only the batteries are to be charged (instrument is ready for
operation in SETUP position), please ensure that the test in-
strument cannot be used by unauthorized personnel during
the charging process.
Measurement may only be performed on voltage-free equip-
ment and devices.
Measurements may not be performed under moist ambient
conditions, where condensation has occurred, or in explosive
atmospheres.
Do not disconnect the measurement cables until the device
under test has been fully discharged.
Observe accident prevention regulation BGV A3, “Electrical
Systems and Equipment”.
Do not perform measurements alone: always have a second
person help if possible.
Make sure that measurement cables are connected correctly.
Do not touch the test probes at any point beyond the finger
guard.
The MERISO PRIME+ may not be used:
If visible damage is apparent
With damaged connector, measuring or test cables (test ca-
bles may not, under any circumstances, be mechanically dam-
aged or kinked, because impaired insulation characteristics
may otherwise result)
If these no longer function flawlessly
If the signal lamps on the PROFITEST204HP/2.5kV
(feature B1) and PROFITEST204HV/5.4kV (feature B2) high-
voltage modules are defective
In such cases, the devices must be removed from service and
secured against any possible inadvertent use.
Send the instrument to our Repair and Replacement Parts service
department in such cases (see chapter 12 on page 67).
GMC-I Messtechnik GmbH 11
Opening of Equipment / Repair
The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid. Even original spare parts may be
installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
2.1 Checklist for High-Voltage Tests
Attention!
!
Measurements may not be performed under moist ambi-
ent conditions, where condensation has occurred or in
explosive atmospheres.
Personnel Safety
Ð Turn the machine under test off and lock it in the off position.
Ð Perform insulation resistance measurement.
Ð Make sure that the system is grounded.
Ð Secure the danger zone with barriers, closing even narrow en-
trance ways.
Ð Set up warning signs such that they are plainly visible.
Ð Set up warning lamps such that they are plainly visible.
Ð Attach the emergency stop button in a plainly visible fashion.
Ð Warn personnel working in the area of possible danger.
Ð Always turn the test instrument off with the key switch and re-
move the key before leaving the work area.
Safety Precautions for the Machine Under Test (recommended)
Ð
Review the circuit diagrams and make a note of all electrical circuits.
Ð Disconnect the neutral conductor from the mains.
Ð Short circuit each electrical circuit to itself.
Ð Disconnect control circuits with overvoltage arresters, if the ar-
resters would be triggered by utilized test voltage.
Ð Disconnect PELV circuits (no high-voltage testing is required
for these circuits).
Ð Test insulation at each circuit with 1000 V.
(If insulation resistance is OK at 1000 V, no failures should
occur during the high-voltage test.)
Ð Disconnect frequency converters.
Ð Caution in TN systems!
The protective conductor is connected to the neutral conduc-
tor in the junction box. High-voltage is thus applied between
the phase conductors and the neutral conductor.
The neutral conductor must be cut if necessary, because it is
not disconnected from the mains by means of fuses.
Setting Up the Test Instrument
High-Voltage Test
Ð
Test all circuits (conductors) against the protective conductor (all
switches in the mains circuit must be turned on, and testing must be
performed upstream and downstream from all relays and contactors).
Ð Remove all shorting devices after testing has been completed.
Testing Without Short-Circuiting
Ð Separately test all conductors from all circuits against the pro-
tective conductor (the machine could be damaged in the case
of arc-over).
Functions Test
Ð The machine must be tested for correct functioning after the
high-voltage test, especially as regards safety functions.
12 GMC-I Messtechnik GmbH
2.2 Significance of Symbols
2.2.1 Symbols on the Device
The symbols on the device have the following meanings:
Indicates EC conformity
Life endangering for the operator if the wiring dia-
gram and the operating instructions are not
observed!
Warning concerning a source of danger
(attention: observe documentation!)
This device may not be disposed of with the trash. Fur-
ther information regarding the WEEE mark can be
accessed on the Internet at www.gossenmetrawatt.com
by entering the search term ’WEEE’.
Report functions
Data management
2.2.2 Symbols in the Operating Instructions
Life endangering for the operator if instructions
identified with this symbol are not observed
Danger for the operator and the device if instruc-
tions identified with this symbol are not observed
2.2.3 Symbols in the User Interface
High-voltage warning,
Voltage at the test probe
between 1000 and 5000 V
Warning concerning a source of danger,
Voltage at the test probe
between 25 and 1000 V
Insulation measurement:
interference voltage at the test probes
All-clear after an insulation test:
voltage at test probes less than 25 V
Timer function: symbolizes test duration
This symbol appears after the voltage measure-
ment value has been frozen (see chapter 4.6 on
page 46).
Indicates progress of the test.
Indicates storage of the measurement value.
!
!
GMC-I Messtechnik GmbH 13
2.2.4 Symbols in the User Interface with Feature B1/B2
High-voltage module ready to be switched
on
Life endangering high voltage of up to 5.35 kV
is present at the test probes of the high-
voltage module.
Test completed successfully,
residual voltage less than 25 V
Display of phase angle
The specified limit value I
MAX
has been exceeded.
Current limiting has switched the device to the
“standby” mode.
The last voltage value measured before interrupting current
was exceeded, as well as interrupting current limit value
I
MAX
are displayed
Excessive temperature inside the test instrument
(high-voltage test)
Allow the device to cool for approximately 10 minutes.
2.3 Special Safety Precautions and Instructions
for Feature B1/B2
Precautions for the prevention of unauthorized start-up:
•Key switch
Precautions for the prevention of inadvertent start-up:
Multiple key operation:
Before test voltage can be applied to the test probes by acti-
vating the trigger at the high-voltage switching pistol, the
START key must be pressed at the primary device.
Two-hand operation:
Each of the high-voltage pistols is equipped with a trigger.
Both triggers must be activated simultaneously in order to ex-
pose both test probes.
High-voltage switching pistol (pistol with yellow marking) with
double safety feature:
If the trigger at the high-voltage switching pistol is pulled to the
first point of mechanical resistance, the test probe is exposed.
High-voltage is not applied to the test probe until the trigger is
pulled beyond this point, assuming the device is switched on.
General Safety Precautions
Integrated signal lamps indicate the switching status of the test
instrument.
Test probes are electrically isolated from the mains:
This prevents current from flowing from the high-voltage
switching pistol to earth.
Current limiting in the event of arc-over:
If the current limit value entered as a device parameter is ex-
ceeded due to arc-over, switching to the “standby” mode
takes place automatically.
•When mains voltage is restored after a power failure, switching to
the “standby” mode takes place automatically.
or
or
14 GMC-I Messtechnik GmbH
Attention!
!
The vent slots in the bottom of the high-voltage module
may not be covered!
Attention!
!
Observe the Rules for Electrical Testing Equipment issued by
the G
ERMAN TRADE ASSOCIATION FOR PRECISION ENGINEER-
ING AND ELECTRICAL ENGINEERING (see enclosed bro-
chure).
Attention!
!
Observe DIN VDE 0104 regulations:
“Setup and Operation of Electrical Test Equipment”
Attention!
!
If safety test probes are used, the operator must inspect
them to assure that the test probes and their cables are
in flawless condition before placing them into service.
All equipment must be inspected for visible, external
damage before use (see chapter 10.3 on page 61
through chapter 10.5 on page 62).
Attention!
!
Before initializing any tests, and before enabling test equip-
ment, make sure that all means of access to the danger
zone are closed, and that all persons have exited the
danger zone.
Attention High-Voltage!
When the trigger at the high-voltage switching pistol is
pulled to the first point of discernible mechanical resis-
tance, the test probe is exposed.
If the trigger is pulled beyond this point of mechanical re-
sistance, high-voltage is applied to the test probe, as-
suming the high-voltage module is in the “standby”
mode.
Attention High-Voltage!
Touch neither the test probe nor the device under test
during the high-voltage test!
Life endangering high-voltage of up to 5.35 kV is present at
the test probes of the high-voltage module!
Limitation of Liability
PCs located in proximity to the test equipment may “crash” in the
event of arc-over, resulting in possible data loss. All data and pro-
grams should be suitably saved, and computers shut down
before performing the high-voltage test. PCs may crash even if no
RS 232 connection has been established.
The manufacturer of the test instrument assumes no liability for
direct or consequential damages to computers or peripheral
devices, or for data loss during performance of the high-voltage
test.
Furthermore, the manufacturer assumes no liability for defects at
devices under test resulting from the high-voltage test. This
applies in particular to electronic components included in the
device under test.
Observe checklist for performance of high-voltage tests in chapter 2.1.
GMC-I Messtechnik GmbH 15
3 Initial Start-Up
The MERISO PRIME+ test instrument is supplied with a base
which serves as a compartment for storing test cables, the guard
cable (feature G1), the test probes and the battery pack
(feature C1).
Attention!
!
The test instrument may only be operated if the
base is properly attached and secured.
3.1 Battery Operation, Feature C1 (not with B1/B2)
The battery pack is required for operation of the MERISO
PRIME+ without mains power. The battery back includes 8 NiMH
batteries. No other battery packs may be used.
The battery pack is charged automatically by the integrated mains
power pack. Optional use of an automotive charging adapter is
also possible. The device must be switched on in this case (any
switch position other than OFF).
Attention!
!
If only the batteries are to be charged (instrument is
ready for operation in SETUP position), please ensure
that the test instrument cannot be used by unauthorized
personnel during the charging process.
3.1.1 Removing the Device Base and Inserting or Replacing
the Battery Pack
Please observe instructions included in this chapter if the device
is to be retrofitted with a battery pack, or if the battery pack
requires replacement.
Attention!
!
Before dismantling the device base:
– Disconnect the device from the measuring circuit at all
poles.
– Disconnect the device from mains power.
– Turn the device off (set the switch to OFF).
Ð Close the cover and position the handle at the side of the de-
vice.
Ð Turn the device upside down, i.e. place it onto its cover.
Ð Unwind the test cables from the cable uptake at the side.
Ð Remove the guard cable from the device base if one is in-
cluded (feature G1).
Ð Loosen the three Phillips head screws at the bottom of the de-
vice base.
Ð Lift the device base up, and set it down at the side of the de-
vice with the feet at the bottom.
Ð Place the battery pack onto the two rubber cushions in the re-
cess in the device base such that the connector cable with
plug connector from the battery pack is on the same side as
the device’s connector cable with socket connector.
Double-check: The opening at the side of the base coincides
with the device’s connection panel.
16 GMC-I Messtechnik GmbH
Ð Connect the two cables to each other and secure the connec-
tion with the snap hooks.
Ð Position the cable such that is cannot be pinched during sub-
sequent assembly.
Ð Carefully set the device onto the base as shown in the photo
by grasping the handle with one hand, and guiding the bottom
of the device into position with the other hand.
Ð Grasp the base and the device, and turn the entire unit upside
down.
Ð Secure the base with the included screws and washers.
GMC-I Messtechnik GmbH 17
Attention!
!
Use only the screws furnished with the device together
with washers. Longer screws may reduce creepage dis-
tances or damage the device.
If a battery pack (feature C1) has been installed and the device is
not connected to the mains, the battery symbol appears at
the display when the device is switched on, and the number of
dark bars displayed in the symbol indicates the battery charge
level.
After installing the battery pack, the device is automatically sup-
plied with power from the batteries.
3.2 Initial Start-Up with Feature B1/B2
3.2.1 Retrofitting Feature B1/B2
Retrofitting with a high-voltage module (feature B1 or B2) can only
be performed by GMC-I Service GmbH.
The MERISO PRIME+ serves to determine the insulation value in
a non-destructive manner. If it is too low, it is possible to detect
the weak point with the high-voltage module (B1 or B2).
3.2.2 Key Switch (feature B1 or B2)
The key switch prevents unauthorized start-up of the high-voltage
module. Keep the key in a safe place which can only be accessed
by authorized personnel.
Torn the key to the “0” position and remove after each test has
been completed.
Note
When the key is removed or the high-voltage module
is turned off, the menu functions for the high-voltage
module cannot be accessed at the
MERISO PRIME+!
Possible Danger of
Pinching!
18 GMC-I Messtechnik GmbH
3.2.3 Indication of Operating States
Signal Lamps at the High-Voltage Module
The signal lamps integrated into the high-voltage module indicate
two different operating states:
Green: Test instrument ready for use
Key switch is in the “I” position (ON).
Power supplies for signal and control circuits at the
high-voltage module are switched on.
Test voltage supply circuits are still switched off, and are still
secured against inadvertent activation.
Attention!
!
All safety precautions required before entering the dan-
ger zone must now have been implemented, amongst
others: set up WS1 warning signs and ZS2 auxiliary
signs in accordance with DIN 40008, part 3.
Red: Instrument ready for operation, caution: danger!
The menu for starting the voltage test has been opened, and
the START key has been activated.
The power supply circuit for the safety test probe is still
switched off, assuming the trigger has not been pulled at the
high-voltage pistol.
The test probes are secured against inadvertent contact, as-
suming the triggers at the high-voltage pistols have not been
pulled.
Attention!
!
All means of access to the danger zone must be closed
when the high-voltage module is in the “standby” mode!
The device must be sent to GMC-I Service GmbH in order to
replace lamps.
External Signal Lamps (feature F1)
The external signal lamps indicate the same operating states as
the signal lamps integrated into the high-voltage module. They
provide additional security at the test site, and must be plainly vis-
ible beyond the boundaries of the danger zone.
The external signal lamps can only be connected to the 12 V/5 W
output jack at the high-voltage module (feature B1 or B2).
Note
For reasons of safety, only Z504D signal lamps available
from GMC-I Messtechnik GmbH may be used.
Refer to chapter 10.7 on page 62 regarding lamp replacement
GMC-I Messtechnik GmbH 19
3.3 Mounting the MERISO PRIME+ with Feature B1/B2 to the
Trolley (feature D1).
Ð Set the complete unit consisting of test instrument and
high-voltage module onto the trolley’s platform (1) such that
the cover of the test instrument can still be opened.
Ð Secure the test instrument’s handle to the vertical supports
with the handle clamps (3). Loosen the screws in the handle
clamps (9) to this end with a 4 mm hollow hex wrench.
Ð Push the handle clamps (3) over the handle and retighten the
screws (9).
Ð Wind the two cables from the safety test probes onto the bot-
tom cable uptake (2 at bottom).
Ð Then wind the two cables from the test instrument’s test
probes onto the top cable uptake (2 at top).
Ð Secure the cables and the test probes with the included rub-
ber straps (12).
Attention!
!
Observe regulations for portable test equipment in ac-
cordance with DIN VDE 0104 (3.6).
Note
Wind the test cables from the safety test probes onto the
provided cable uptakes at the back of the trolley after
each use. Test cables may not, under any circumstances,
be mechanically stressed or kinked, because impaired
insulation characteristics may otherwise result. Do not
forget that externally caused damage is not always rec-
ognizable.
Attention!
!
Inspect the test cables for mechanical damage before
placing the high-voltage module into service each time it
is used.
10
11
8
2
3
2
5
6/7
1
4
9
Item Designation Qty. Order Number
1 Base 1 3-117-193-01
2 Cable uptake 2 3-326-653-01
3 Handle clamp 2 3-326-652-01
4 Cover 2 3-164-609-01
5 Wheel 2 3-419-038-01
6 Washer 4 3-740-013-01
7
Retaining ring
2 3-743-024-59
8 Screw 6 3-712-007-10
9 Hex nut 6 3-730-119-12
10 Screw 2 3-716-018-24
11 Trolley frame 1 3-121-111-01
12 Rubber strap 2 3-326-627-01
13
Protective cover
1 3-171-302-01
Caddy 204 Trolley
20 GMC-I Messtechnik GmbH
3.4 Mains Connection
Ð Connect the test instrument to 230 V or 120 V mains power
(depending on the country-specific design) with the included
mains connector cable.
Attention!
!
If a connection via earthing contact outlet is not possible,
disconnect the mains from all sources of voltage before
connecting the cables of the coupling socket via pick-up
clips with the mains connections, as shown in the draw-
ing.
If a battery pack has not been installed, the mains connection
symbol appears after the test instrument has been
switched on.
If a battery pack has been installed (feature C1), it is automatically
recharged. The battery symbol appears at the display, and
the blinking bars shown in this symbol indicate the remaining
capacity which still needs to be charged. The device need not be
disconnected from mains power after the batteries have been fully
charged (overcharge protection).
3.4.1 DC Operation (automotive charging adapter in preparation)
If a battery pack (feature C1) has been installed and the device is
not connected to the mains, the battery symbol appears at
the display when the device is switched on, and the number of
dark bars displayed in the symbol indicates the battery charge
level.
After installing the battery pack, the device is automatically sup-
plied with power from the batteries.
The automotive charging adapter (feature J1) can be used to
operate the device and charge the batteries.
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
U
L–N
= 230 V
GMC-I Messtechnik GmbH 21
3.5 User Interface
Measuring and testing with the MERISO PRIME+ is quick and
easy. The integrated user interface provides the operator with in-
formation regarding required operating steps, operator errors,
measurement results and more for all measuring functions. All in-
formation and measurement results appear at a dot matrix LCD
panel in plain text.
The integrated user interface is entirely adequate for the great
majority of all tests and measurements. Nevertheless, operators
should carefully read and understand the contents of these oper-
ating instructions.
3.6 Help Functions
Help texts can be queried and displayed at the LCD for all mea-
suring and test functions, and for practically all settings.
Example: Display Contrast Function
SETUP
Switch Position
Ð Press the STORE key in order to query online help:
STORE
Query Help
Ð Press the same key once again, or the MENU key, in order to
exit the help function:
MENU
Exit Help
or
Ð Start the selected function directly from the help menu by
pressing the START key:
START
Start Function
22 GMC-I Messtechnik GmbH
3.7 Setup
Various basic settings can be selected for the test instrument with
the selector switch in the SETUP position.
SETUP
Select Parameter
MENU
Open Menu for Entering Parameters
All setting and changes entered to the menu accessed in the
SETUP switch position are saved to memory automatically, and
are retained after disconnecting the device from mains power.
MENU
Exit Function
Restoring default setting – standard values
Switch off the test instrument: switch position OFF.
Then press and hold the MENU and i/STORE keys simultane-
ously, and turn the function key to TEST position. The standard
values of the default setting are thus restored.
3.7.1 Adjusting Contrast and LCD Illumination
Contrast can be increased or decreased with this function.
LCD illumination can be switched on or off as well.
SETUP
Select Display Contrast
MENU
Open Entry Menu
Adjust Contrast
START
Switch Illumination On or Off *
* Background illumination is switched off automatically after approximately 2.5 min-
utes if none of the keys are activated during this time. Each time a key is activated,
background illumination is extended for an additional 2.5 minutes.
GMC-I Messtechnik GmbH 23
3.7.2 Setting Date and Time
The test instrument’s internal clock can be set with this function.
The clock continues to run after supply power has been discon-
nected. Date and time are included in the test reports.
Entries are made using the following formats:
Date: DD.MM.YYYY (Day.Month.Year)
Time: hh:mm (Hour:Minute)
SETUP
Select Date/Time
MENU
Open Entry Menu
(and select date or time)
START
Select Cursor Position
Change Values
24 GMC-I Messtechnik GmbH
3.7.3 Beeper Settings
The beeper can be completely deactivated or activated as indi-
cated in the table below.
* a good measurement result (passed) is indicated with one long acoustic signal,
a bad result (failed) with three brief acoustic signals.
SETUP
Select Beeper
MENU
Open Entry Menu
Select Beeper Function
Beeper
deactivated
Beeper
activated
Beeper for
measuring functions
Beeper for
high-voltage
Beeper
upon key stroke
——
Beeper upon good/
bad measurement
results *
••
Intermittent
beeper for high-
voltage > 1000 V
••
Intermittent beeper
for operation with
Feature B1/B2
••
GMC-I Messtechnik GmbH 25
3.7.4 Executing the Self-Test
This function displays important device data first. If required, dis-
play lamp and relay functions, as well as other options, can be
tested automatically and results appear at the display.
SETUP
Select Self-Test
MENU
Open Entry Menu
LCD Test
Neighboring horizontal and vertical segments of the LCD panel
are activated separately or test images are indicated by means of
the following tests. If one of the segments fails to blink, send the
test instrument to an authorized service center for repair.
START
Battery Voltage and LED Test
START
Continue Test
Results: OK
Error message:
– Test failed
– component defective
– component not available (e.g. batteries not inserted or
high-voltage module not available)
Note
Before a check mark appears to indicate correct func-
tioning the LEDs must first blink, and the relays must
switch.
3
26 GMC-I Messtechnik GmbH
3.7.5 Downloading a Language or a Software Update
If you require a language other than one of those included with
the instrument, it can be downloaded with the help of WinProfi*
PC software. The data file with the desired language is transmit-
ted to the test instrument via the serial interface. The previously
installed language is overwritten.
Software updates can be downloaded with the help of the same
program, regardless of the currently installed language.
Note
You can use WinProfi software at the PC side (included
with the MERISO PRIME+) to download a language to
the test instrument. This software includes all of the func-
tions required for communications between the MERISO
PRIME+ and the PC. A description of the program is
included in the online users manual which can be
accessed from WinProfi.
* WinProfi is currently applicable up to Windows 7 (32 Bit)
A Installing WinProfi to the PC and Starting the Program
Ð Download the WinProfi software from our homepage:
http://www.gossenmetrawatt.com
(
Products
Software
Software for Testers
WinProfi)
Ð Unzip the zip file „winprofi.zip“.
Ð Install the software on your PC by executing file
Setup_WinProfi_Vx.xx.exe.
Ð Select the desired language for the WinProfi software and for
the user interface of the test instrument.
Ð Follow the instructions which appear at the monitor.
The program is added to your start menu after installation.
Displaying and printing out the online users manual
(contains information concerning the software which is not
included in these operating instructions).
GMC-I Messtechnik GmbH 27
B Prerequisites for Software Update or Exchange of Data
Ð Find the interface to which the MERISO PRIME+ is con-
nected.
Note
Always start this function first, before performing an update
or changing report templates. After starting this function,
WinProfi loads the report files required specifically for the
connected instrument.
Due to the fact that WinProfi has been created for use
with several types of test instruments, incorrect test
reports may otherwise be loaded, or erroneous options
may be made available.
Ð Querying information regarding current software version
C Transmitting a Software Update to the Test Instrument
Ð PC: Select Update all from the update menu. Follow the instruc-
tions which appear at the monitor.
Depending upon the utilized PC, updating takes from 1 to 2 min-
utes.
The green LED indicates that the instrument is ready to receive
data. If the PC and the test instrument are correctly synchronized,
the yellow LED lights up and the green LED goes out. The red
LED lights up during programming sequences. After transmission
has been completed, the LED goes out and the device is
restarted.
The message “Transmission done” appears at the PC.
Attention!
!
The instrument may not, under any circumstances, be
switched off during transmission, nor may the connec-
tion between the instrument and the PC be interrupted!
Subsequently select “Language” (loading the respective lan-
guage), if applicable.
28 GMC-I Messtechnik GmbH
D Organizing Report Data
Ð Connect the test instrument directly via cable Z3241 with the
PC.
Note
Do not connect the PC to the PSI module.
Sending and receiving data
Printing out data
Editing and transmitting report templates
6
3
5
2
1
4
GMC-I Messtechnik GmbH 29
4 Performing Tests
Each of the following tests is automatically assigned to one cur-
rently available object number. Any given test which is saved to
memory with the “STORE” key is stored with reference to this
number. The object number can be entered in the “DUT data”
menu (see chapter 5.1 on page 48).
Measuring mode operation is always carried out according to the follow-
ing sequence:
Select the desired test.
Enter test parameters (if required).
Start the measurement or return to the initial menu.
Save measured values to memory and enter a description
(if desired).
Observe the following for all tests:
Safety features and precautions, see chapter 2.
Disconnect the device under test from all sources of voltage
first.
Test for absence of voltage.
Do not disconnect the measurement cables until the device
under test has been fully discharged.
4.1 Insulation (resistance) Test
Test voltage can be set between 100 V and 5,000 V for this test.
This voltage, as well as the allowable limit value for insulation
resistance, can be selected as a test parameter.
Voltage at the device under test appears at the LCD in addition to
measured insulation resistance.
Discharge
After the measurement has been completed, the device under
test is discharged automatically in order to assure zero voltage.
The falling voltage value can be observed at the display during the
discharging process. The measured insulation resistance value
remains in memory during this period of time.
Interference Voltage
If voltage is present at the test probes after mea-
surement is started, an “external voltage warning
appears at the display. The level of the interference
voltage is shown at the display. The insulation test is aborted after
the test probes have been removed from the DUT. The test has to
be conducted again on the de-energized DUT.
30 GMC-I Messtechnik GmbH
Attention!
!
Do not touch the test probes during measurement!
A direct voltage of up to 5000 V is present between the
test probes.
Note
Testing the Measurement Cables:
Initial testing should be performed after short-circuiting
the measurement cables with their test probes. The
instrument should display nearly zero MΩ. A broken mea-
surement cable can be detected in this way.
TEST
Select the Test
MENU
Open the Submenu
4.1.1 Setting Test Parameters
The type of utilized test voltage can be selected here:
Fixed test voltage
Variable test voltage
(e.g. lower test voltage for voltage-sensitive components)
Test voltage in a sequence of steps (U
STEP
= step voltage), e.g. in
order to allow for better evaluation of insulation quality with re-
gard to dielectric strength
Additional parameters must be set as well depending upon the
selected type of test voltage.
MENU
Select Parameter
Set Value
GMC-I Messtechnik GmbH 31
Significance of Parameters with Variable and Fixed Test Voltage
Test voltage U
nom
: Test voltage in fixed steps
(100/250/500 V, 1.00/1.50/2.00/2.50/5.00 kV)
U
var
: Variable test voltage
(100 V 5.00 kV in 50 V steps)
U
ins
Test voltage
Limit R
ins
Limit value for insulation resistance
(If value stays below limit value: measurement failed)
Test
duration Maximum test duration in seconds
<Auto>: measurement until steady state is reached
Time in sec.: Measurement = 1 to 120 seconds
>>>>: cont. measurement: stop with MENU key
Time in min log:Measurement = 2 to 60 minutes with
automatic storage of 30 measured values at equal time
intervals including information on the current test time.
(see chapter 4.1.4 on page 32).
Significance of Parameters with Step Voltage
Test v. U
step
: Test voltage as a sequence of steps
U
(INS) END
Final value of the step sequence
(250 V 5.00 kV in steps of 250 V)
ΔR
Lim
Maximum allowable statistical deviation from the mea-
sured mean value
Δt/250V Dwell time per step (10 s 50 s)
Resetting parameter values to default setting, see page 22.
Note
In the event of very high insulation resistances the capacitive
influence of the person performing the measurement and/or of
the measuring cable may distort the measured value.
4.1.2 Starting the Test (U
NOM
, U
VAR
)
START
Display of results
after completion of test
4.1.3 Starting the Test (U
STEP
)
START
32 GMC-I Messtechnik GmbH
Statistical analysis
Press the CURSOR DOWN or CURSOR UP keys to return to the previous
LCD display.
Note
If the measuring range is exceeded (overflow of display
range), no statistical analysis will be performed. In this
case, the individual measured values are listed, but no
value for ΔR
MAX
is calculated.
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
4.1.4 Data logger function (min log)(as of firmware version AH)
MENU
Select Parameter
If you choose time parameter setting "min log" from the menu,
exactly 30 measured values are automatically stored during the
selected test duration at time intervals between 4 seconds and
2 minutes. This makes data logger operation possible for measur-
ing durations between 2 and 60 minutes. In this case, the residual
time of the current measurement is not indicated in seconds, but
in minutes. While the test is being performed, icon “Stored” is
briefly shown in the display and the yellow signal lamp “Testing ...
disappears briefly each time a measured value is being stored. If
the signal generator is activated, an acoustic signal sounds addi-
tionally.
In this function, the current test time is stored for each measure-
ment instead of the parameter for the test duration to ensure the
temporal correlation between measured value and time of mea-
surement.
Note
Before the start of each test with data logger as
described in chapter 5.1, select a new device under test
to avoid the measured values being stored successively
within the same DUT.
This considerably facilitates the subsequent evaluation of
the stored measured values.
GMC-I Messtechnik GmbH 33
4.1.5 Measurements with the Guard Cable (feature G1)
The measurement of very high resistances necessitates extremely
minimal measuring current and may be rendered problematic as a
result of influences such as electromagnetic fields, humidity or
surface current. An accurate test set-up is thus absolutely essen-
tial.
A guard cable must be used for measurements within a range of
100 GΩ (10 GΩ) 1TΩ, in order to prevent surface current from
distorting measurement results. The guard rings prevent current
at the surface of the insulation material from flowing from the
+measurement cable to the measurement cable, instead of
through the insulation material itself.
Ð Insert the plug from the guard cable into the appropriate jack in
the test instrument.
Ð Plug the alligator clip onto the guard cable test probe.
Ð Connect the alligator clip to the guard ring between the two
measuring points at the insulation material under test.
Ð Refer to chapter 4.1.2 on page 31 regarding the measuring se-
quence
Note
The following materials can be used as guard rings: alu-
minum foil, copper foil or metallic hose clamps.
+Measurement Cable
Measurement Cable
Guard Cable
Conductor
Insulation
Material
Guard Rings
Contact Ring
34 GMC-I Messtechnik GmbH
4.2 High-Voltage Test with Feature B1/B2
The electrical equipment of the machine under test must
withstand a test voltage of twice its own rated voltage value or
1000 V~ (whichever is greater) applied between the conductors of
all circuits and the protective conductor system for a period of at
least 1 second. The test voltage must have a frequency of 50 Hz,
and must be generated by a transformer with a minimum power
rating of 500 VA.
Note
Feature B must be included for testing in accordance
with DIN VDE 0113.
Feature B2 is used for testing at up to 5 kV.
The MERISO PRIME+ automatically recognizes which type of high-voltage
module has been connected. Communications between the test instrument and
the high-voltage module can be tested as described in chapter 3.7.4.
Attention!
!
Before beginning work, the test instrument, the high
voltage cables and the high-voltage pistols must be
inspected to assure that they are in flawless condition
(see also chapter 10.5 on page 62).
4.2.1 Performance Test (preparation for testing)
Execute the following performance test in the specified order from
a) through e).
a) Key Switch
Ð Set the key switch to the “0” position.
Neither the green nor the red signal lamps may light up.
b) Voltage Test
Ð Set the key switch to the “I” position.
The green signal lamps must light up.
Ð Select the insulation test with the
or the key.
TEST
Select the Test
MENU
Open the Submenu
The test voltage parameter is switched from U
NOM
or U
STEP
to
AC-HV.
Ð Set parameters for test duration, test voltage, interrupting cur-
rent I
MAX
and rise time to the desired values (see chapter 4.2.2
on page 36).
Note
Do not set I
MAX
to pulse control mode!
Ð Press the START key.
The red signal lamps must light up.
The following values appear at the LCD:
U
T
= –––V
I
T
=–––mA
The following symbols appear:
GMC-I Messtechnik GmbH 35
Attention High-Voltage!
Touc h neither the test probes nor the device under test
during the high-voltage test!
Life endangering high-voltage of up to 5.3 kV is present at the
test probes of the high-voltage module!
Ð Pull the trigger of the identified high-voltage switching pistol to
the limit stop and hold.
The following values and symbols appear at the LCD:
U
T
= acquired voltage value
I
T
00.0 mA or 0.00 mA
After the specified test time has elapsed, test voltage is
switched off automatically and the
following appears:
If the trigger (switch) is release prematurely, the test is aborted.
c) Shutdown Functions
Ð Short circuit the two high-voltage pistols.
Ð Pull the trigger of the identified high-voltage switching pistol to
the limit stop and hold.
The instrument is shut down immediately.
The green signal lamps must light up, and the red signal lamps
may not light up.
The following values appear at the LCD:
U
T
= XXX V
I
T
= > XXX mA
The last voltage value measured before interrupting current was exceeded
and breaking current limit value I
MAX
are displayed.
The following symbols appear:
d) Power Failure
Ð Press the START key.
The red signal lamps must light up.
Ð Set the key switch to the “0” position and wait at least 1 sec-
ond.
Ð Set the key switch to the “I” position.
The green signal lamps must light up, and the red signal lamps
may not light up.
e) Indicator Signals
Signal Lamps
Two lamps are included on the test instrument, arranged
diagonally, for the indication of operating states. If both “ready for
operation” lamps (red) go out, test voltage is no longer generated.
Do not start voltage testing unless all signal lamps indicate correct
operation.
Replace defective lamps as described in chapter 10.6 under
“Maintenance”.
Acoustic Signals (acoustic signal “on” or “measuring mode operation”)
Refer to chapter 3.7.3 on page 24 regarding activation.
0
t [s]
1
2
3
0
t [s]
1
2
3
0
t [s]
1
2
3
Ready
Test in Progress
I
T
>I
MAX
Repetitive sequence
Single sequence
Repetitive sequence
36 GMC-I Messtechnik GmbH
4.2.2 Setting Test Parameters for High-Voltage Testing
The following parameters can be set in the “standby” mode with
illuminated green signal lamps:
Test v. AC-HV
U
AC.MAX.
Max. test voltage, final value after T
R
has elapsed
Setting limits: HP-2.5kV: 250 V 2.50 kV
HV-5.4kV: 650 V 5.35 kV
I
MAX.
: Max. allowable current before
high-voltage is switched off
Setting limits: HP-2.5kV: 10.0 250 mA
HV-5.4kV: 0.50 10.0 mA
The pulse mode symbol >>>>>>> can be selected with the
key at less than 10.0 or 0.50 mA.
T
D
: Test duration at maximum test voltage U
AC.MAX.
(not including rise time T
R
)
Setting limits: 1.00 120 s, the continuous
operation symbol >>>>>>> can
be selected at under 1.00 s
with the key.
T
R
: Rise time: time, during which test voltage
rises to the selected value U
AC.MAX.
Setting limits: 100 ms 99.9 s
MENU
Select Parameter
Set Value
Test
voltage
Rise
Test duration Decline*
Tes t pe r io d
Time
* constant = 0.1 s
U
T
T
R
T
D
GMC-I Messtechnik GmbH 37
4.2.3 High-Voltage Testing Sequence
Attention!
!
Before initializing any tests, and before enabling the test
equipment, make sure that all means of access to the
danger zone are closed, and that all persons have exited
the danger zone.
Ð Set the key switch to the “I” position.
The test instrument is switched to the “standby” mode.
The green signal lamps are illuminated.
Ð Check the test parameters.
Ð Press the START key.
The test instrument is switched from the “standby” mode to
“ready for operation”. The red signal lamps are illuminated.
START
Attention High-Voltage!
Touc h neither the test probes nor the device under test
during the high-voltage test!
Life endangering high voltage of up to 5.3 kV is present
at the test probes of the high-voltage module!
Ð Position the high-voltage pistols in close proximity to the de-
vice under test.
Ð Pull both triggers, but only pull the trigger at the high-voltage
switching pistol up to the first point of mechanical resistance.
The test probes are enabled.
Ð Establish contact with the electrical circuits.
Ð Pull the trigger of the high-
voltage switching pistol to
the limit stop.
High-voltage is now switched
to the test probes.
Test time is indicated by an
internal beeper which can be
activated or deactivated, and
visually by the timer symbol
(ton = t
R
+ t
D
).
The test continues until test
time (selected rise time plus
test duration) has elapsed:
ton = 000.
Ð For subsequent testing of neighboring circuits, allow the trigger
to return to the first point of mechanical resistance, establish
contact with the next circuit and pull the trigger to the limit stop
once again. Testing is started anew.
38 GMC-I Messtechnik GmbH
Saving Measured Values
After the test sequence has been completed, the last measured
U
T
and I
T
values remain at the display. Results can be saved to
memory with or without a comment. An automatically generated
consecutive number is assigned to the results for allocation to the
respective test.
A description with up to 15 characters can be entered for the
measurement to the field at the bottom left. See chapter 5.1.1 on
page 49 regarding entry.
The measured values at the display are overwritten when the next
test is executed.
STORE
Press briefly: Save results
Press & hold: Save results and initialize entry field
In the continuous operation mode (test duration set to >>>>>>>),
no values are saved to memory for U
T
and I
T
.
Aborting the High-Voltage Test Before Completion
The high-voltage test can be aborted prematurely at any time by
releasing the trigger at the high-voltage switching pistol.
If testing is aborted by releasing the trigger at the high-voltage
switching pistol before the selected test duration has elapsed,
displayed U
T
and I
T
values may deviate from the last measured
values as a result.
Note
Breakdown Voltage
If interrupting current I
MAX
is exceeded
before the selected test voltage has been
reached, test voltage U
T
and current I
MAX
measured at
this point in time appear at the display preceded by the
“>” symbol, and are saved to memory. The appropriate
symbol is displayed simultaneously.
GMC-I Messtechnik GmbH 39
4.2.4 Pulse Control Mode
We recommend selecting the pulse control mode for
troubleshooting purposes (arc-over pinpointing).
Interrupting current I
MAX
is fixed to a value of approximately
125 mA with feature B1, and to approximately 5 mA with feature
B2 in the pulse control mode. If this value is exceeded, test
voltage is switched off within 0.5 ms. After approximately 0.6 s,
test voltage is increased from 0 to its final value within a period of
0.2 s, or is switched back off if interrupting current is exceeded.
Ð Set I
MAX
to >>>>>>> in order to activate the pulse control
mode (see chapter 4.2.2 on page 36).
Ð If continuous operation is required, set test duration to >>>>>>>
as well (see chapter 4.2.2 on page 36).
4.2.5 Ending the High-voltage Test
Ð Release the triggers at the high-voltage pistols.
Ð Press the MENU key.
Ð The signal lamps change from red to green.
Ð The test system must be removed from service (signal lamps
off) if left unattended.
Ð Set the key switch to the “0” position and remove the key, and
secure the device against unauthorized use.
Test breakRise
U
T
t
0,6 s
0,2 s
40 GMC-I Messtechnik GmbH
4.3 Polarization Index Measurement
A polarization index test is recommended for electrical machines.
This procedure involves expanded testing of insulation resistance.
DC measuring voltage from the MERISO PRIME+ is applied to
the insulation for a duration of 10 minutes. Measured values are
documented after one minute, and after ten minutes. If the insula-
tion is good, the value measured after ten minutes is higher than
the value measured after one minute. The relationship between
the two measurement values is the polarization index. Charged
material within the insulation is aligned due to the application of
measuring voltage over a long period of time, resulting in polariza-
tion. The polarization index indicates whether or not the charged
material contained in the insulation can still be moved, thus allow-
ing for polarization. This, in turn, is an indication of the condition of
the insulation.
The following rules apply in general:
PI values < 1 Troubleshooting is required.
PI values = 1 2 Maintenance is recommended.
PI values = 2 4 DUT is OK, no immediate action is
required. Preventive maintenance can be
planned according to the number of tests to
be performed.
PI values > 4 DUT is in flawless condition.
Applications
Determination of moisture and contamination levels
Absorption Index (DAR)
– DC Charging Test
The absorption index test is part of the polarization index test.
Insulation resistance measurements are placed in relationship to
one another after 30 and 60 seconds.
Application: Faster version of the polarization index text
TEST
Select the Test
MENU
Open the Submenu
GMC-I Messtechnik GmbH 41
4.3.1 Setting Test Parameters
MENU
Select Parameter
Set Value
Significance of Parameters
U
ins[nom]
Nominal insulation voltage
100/250/500 V, 1.00/1.50/2.00/2.50/5.00 kV
PI(t2/t1) Polarization index PI is the relationship between insula-
tion resistance values measured after 1 minute PI(t1),
and after 10 minutes PI(t2). Other durations can be
selected if required.
PI (limit) Selected polarization index limit value: appears in the
footer in the measuring menu
DAR(t2/t1) Absorption ratio DAR is the relationship between insu-
lation resistance values measured after 30 seconds
DAR(t1), and after 60 seconds DAR(t2). Other dura-
tions can be selected if required.
DAR (limit) Selected absorption ratio limit value: appears in the
footer in the measuring menu
For resetting parameter values to default setting, see page 22.
4.3.2 Starting the Test
START
The two countdown timers (10:00 and 02:00) display remaining
test time for each respective function.
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
42 GMC-I Messtechnik GmbH
4.4 Breakdown Voltage
Step-Voltage Test – DC Charging Test
Voltage is increased up to the limit value (selected maximum test
voltage) over the selected rise time during the breakdown voltage
test. If voltage breakdown occurs before this value is reached,
breakdown voltage U
brk
is displayed.
If the selected test voltage is reached, testing is continued at this
voltage level until the time period selected under “test duration”
has elapsed.
If no breakdown occurs before the selected test duration has
elapsed, the test has been passed, in which case the following
display appears: U
brk
– – –. Breakdown voltage is displayed in all
other cases.
Applications
For the recognition of cracks and holes in the insulation
TEST
Select the Test
MENU
Open the Submenu
GMC-I Messtechnik GmbH 43
4.4.1 Setting Test Parameters
Set Parameters
Significance of Parameters with Variable and Fixed Test Voltage
Test voltage U
nom
: Test voltage in fixed steps
(100/250/500 V, 1.00/1.50/2.00/2.50/5.00 kV)
U
var
: Variable test voltage (100 V 5.00 kV)
U
ins
Test voltage (see fixed steps above)
Rise Time after which maximum test voltage will be reached
(5 s 300 s)
Test
duration Test duration at maximum test voltage
(AUTO, 1 s 120 s,
>>>> (continuous, burn mode))
For resetting parameter values to default setting, see page 22.
4.4.2 Starting the Test
START
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
44 GMC-I Messtechnik GmbH
4.5 Capacitance Measurement and
Determination of Dielectric Discharge (optional)
Capacitance measurement is performed with adjustable test volt-
age by charging with constant current. The measurement is per-
formed with automatic range selection. One of the nominal test
voltages (U
nom
), or a freely selectable test voltage (U
var
) can be
used.
Dielectric Discharge Test (DD)
This test was developed for use with large machines and is per-
formed, as its name indicates, during discharge. It designates the
ratio of the discharge current after one minute and the product of
voltage by capacitance:
DD values > 4: Troubleshooting is required
DD values = 2 4 Maintenance is recommended
DD values < 2 DUT is OK, no immediate
action is required.
Application: Determination of the magnitude of absorbed contami-
nation or moisture
TEST
Select the Test
MENU
Open the Submenu
4.5.1 Setting Test Parameters
Set Parameters
For resetting parameter values to default setting, see page 22.
DD
I
Disch
1min()
UC
------------------------------
1000=
GMC-I Messtechnik GmbH 45
4.5.2 Starting the Test
START
After capacitance has been determined, the instrument automati-
cally continues with dielectric discharge testing. This takes about
30 minutes. However, the measurement can be aborted at any
time by pressing the MENU key. Only capacitance is displayed in
this case.
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
46 GMC-I Messtechnik GmbH
4.6 Voltage Measurement (protection against residual voltage)
EN 60204 requires that voltages in excess of 60 V occurring at all
accessible machine components during operation must drop to a
residual voltage value of 60 V or less within a period of 5 seconds
after supply power has been switched off.
Testing for the absence of voltage is performed by means of a
voltage measurement which involves measuring discharge time
with the MERISO PRIME+. Elapsed time as of disconnection
from supply power is displayed continuously at the LCD during
discharge, until the limit value of 60 is reached.
In addition to voltage type (AC, DC or AC+DC), frequency is mea-
sured and displayed as well. The mean value is always read out.
In addition to voltage and frequency, voltage type (AC, DC or
AC+DC) appears at the symbol field in the display.
If voltage drops by more than 5% of the measured value during a
single measuring period (approx. 0.7 sec.) while performing a volt-
age measurement which results in a value of greater than 60 V, a
time measurement is started automatically in order to test protec-
tion against residual voltage. This time measurement can be
observed at the display in the symbol field. Time measurement is
stopped after voltage drops to below the limit value of 60 V. Time
required to discharge down to 60 V then appears at the display in
seconds. Voltage measurement continues until maximum allow-
able discharge time has elapsed (normally 5 seconds). The results
of the voltage measurement are then “frozen”, and the
DATA HOLD symbol appears at the display. If the 60 V limit value
is not reached before allowable discharge time has elapsed, the
voltage value is “frozen” and time measurement continues until
the 10 second time limit has elapsed. If voltage is still greater than
60 V after 10 seconds, the test is aborted and the instrument
restarts voltage measurement.
When the display is “frozen”, voltage measurement is not started
again until:
The START key is pressed.
The measured value is saved to memory by pressing the
i/Store key.
Voltage measurement is restarted from the menu.
Or voltage at the test probes climbs back up to a value of
greater than 60 V, or if the next measured value after a mea-
surement of greater than 60 V is even higher, i.e. if a climbing
tendency can be observed.
Both elapsed time and the prevailing voltage value after allowable
discharge time has elapsed are saved to memory by pressing the
i/Store key. If no discharge takes place, only the voltage value is
saved.
Note
If, for example, conductors are exposed when a machine
is switched off – e.g. if plug connectors are disengaged –
and if these are not protected against direct contact,
maximum allowable discharge time is 1 second!
TEST
Select the Test
MENU
Open the Submenu
GMC-I Messtechnik GmbH 47
4.6.1 Setting Test Parameters
The discharge time during which the voltage value must drop to a
non-hazardous value of less than 60 V can be specified as a limit
value between 1.0 and 9.00 seconds. The following message
appears in order to indicate that the limit has been reached:
“residual voltage less than 60 V!”.
Set Discharge Time
Test duration is always set to continuous measurement in this case
because residual voltage testing is triggered automatically and
voltage measurement is always active for safety reasons.
For resetting parameter values to default setting, see page 22.
4.6.2 Starting the Test
START
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
48 GMC-I Messtechnik GmbH
5 Processing, Reorganizing and Deleting Data
Previously stored data and data records can be processed with
the selector switch in the data management position.
The following functions are available:
DUT data
An object number can be assigned here, and measurements
can be saved with reference to the selected number. Beyond
this, a description can be entered for the respective object.
Data Processing
Individual data resulting from a measurement (e.g. insulation
measurement) performed on a previously selected object can
be displayed with this function, and subsequently deleted if de-
sired.
Data Reorganization
Previously entered objects can be deleted from memory with
this function, and memory can be tested as well.
Clear Memory
This command deletes data records previously saved to mem-
ory for all objects. Full memory capacity is made available by
means of this procedure.
Memory Occupancy
Current memory occupancy is continuously displayed as a bar
graph for the first three menu items listed above.
Depending upon the number of objects stored to memory
(max. 254), up to 1,600 measurements can be saved.
5.1 Select Device under Test (DUT)
Select a Command
MENU
Submenu
Object no.
Select Number
Status: A filled file cabinet indicates that measurement data
have already been saved to memory for the selected
object.
GMC-I Messtechnik GmbH 49
5.1.1 Entering Descriptions
MENU
Description
Description: A description can be entered here for an object, which
is displayed and subsequently printed out when the respective
object is selected.
There are five different ways to enter descriptions:
Convenient entry with the PSI module (feature I1)
In this case, descriptions are entered using the alphanumeric
keypad included with the PSI module (see operating instruc-
tions for SECUTEST PSI).
Entry with the help of a barcode scanner (accessory no. B3261)
or RFID scanner (accessory no. Z751G)
Entry from a terminal
Entry with PC keyboard via the PROFI-MFII adapter
(accessory no. Z504H)
Entry using keys at the MERISO PRIME+
Select Letter, Number or Character.
STORE
Move Cursor Left
START
Move Cursor Right
MENU
Completed entries must be acknowledged with the MENU key.
The entry is then saved to memory. Alternatively, the “Store” key
can be pressed at the PSI module.
Up to 250 characters can be entered. The entry line scrolls auto-
matically.
A comment can be entered for each test step in the same way,
immediately after the test has been performed. However, the
length of these comments is limited to 15 characters.
50 GMC-I Messtechnik GmbH
5.1.2 Copying Descriptions
Ð Select the object with the description to be copied and press
the MENU key 3 times.
Ð Select the new object and press the MENU key.
The copied text is displayed and can be edited.
5.1.3 Deleting Descriptions
Ð Simultaneously press the i/STORE and START keys.
The description is deleted and the cursor is sent to the first char-
acter entry position in the description field.
5.2 Processing Data (checking measured values)
Measured values can be checked with this function, and individ-
ual measurements can be deleted if desired.
Select a Command
MENU
Select an Object
STORE
Delete the Object
“Data record deleted” appears after deletion of the respective
measurement has been started.
In order to delete a data record, the i/Store key must be pressed
and held for approximately 1 second, thus avoiding inadvertent
deletion. If the key is not released, all subsequent data records
are then deleted sequentially at a rate of 1 record per second.
GMC-I Messtechnik GmbH 51
5.3 Data Reorganization
5.3.1 Deleting Already Printed Data
Data for individual DUTs which have already been printed out in
report form, or DUTs which have been entered although no mea-
sured values have been saved for them yet, can be deleted here if
desired.
Select a Command
MENU
Submenu
START
Start Deletion
STORE
Delete the Object
Security acknowledgement must be entered before deleting each
individual object. As soon as objects whose data have already
been printed out in report form have been deleted, a message to
this effect appears. If the START key is pressed once again, a
print-out is generated including all DUTs for which measured data
are stored to memory.
5.3.2 Memory Test
The memory module is checked for errors with this function.
Errors can be corrected in some cases. Follow the instructions
which appear at the display.
MENU
START
Test Memory (press 3 times)
52 GMC-I Messtechnik GmbH
5.4 Clearing the Memory
This command deletes data records previously saved to memory
for all DUTs. Full memory capacity is made available by means of
this procedure.
Attention!
!
Data should be downloaded to a PC and saved before
clearing the memory!
Select a Command
MENU
Submenu
START
Execute the Command
A security acknowledgement must be entered before memory is
cleared. Final deletion is initialized with the i/STORE key.
STORE
Clearing the memory entirely may takes several seconds. “Mem-
ory cleared” is then displayed.
GMC-I Messtechnik GmbH 53
6 Printing Test Results
The test instrument’s report functions can be activated with the
selector switch in the printer position. The following functions are
available:
Print out values at the PSI module (feature I1):
Measured values for a selected object can be printed out at
the utilized printer module via the serial interface.
Download templates from a PC (see WinProfi PC software in
chapter 3.7.5):
Report layouts can be specified by the user. A template with a
header, a text body and a footer is required to this end. A tem-
plate created by the user at a PC can be downloaded to the
test instrument via the serial interface.
Select Printer
MENU
Printer Menu
Select an Object or a Number
START
Start Printing *
Status: A filled file cabinet indicates that measurement data
have already been saved to memory for the selected
DUT.
A printer symbol appears as long as data is being
transmitted to the printer.
* A print-out which has already been started can only be aborted by turning the se-
lector switch. An error message is generated if the printer does not work. Printing
must be restarted in this case.
54 GMC-I Messtechnik GmbH
7 Characteristic Values
Standards DIN EN 61557-1:2007
DIN EN 61557-2:2008
VDE regulations VDE 0413 Part 1:2007
VDE 0413 Part 2:2008
Insulation Resistance
Test duration: automatic (until measured value stabilizes),
manual (1 ... 120 s) or continuous measurement (lock function)
Polarization Index (PI), Absorption Ratio (DAR)
PI and DAR are calculated values. Insulation measurement specifications apply.
Insulation Test Voltage
Variable test voltage selectable in 50 V steps
Short-circuit current less than 2 mA with test voltage to 1.00 kV
Voltage Measurement
Frequency Measurement
Measured quantity voltage: 50 V ... 1 kV
Breakdown Voltage
Display Range
[Ω]
Measuring
Range [Ω]
Test Voltage Intrinsic Uncer-
tainity
Measuring Uncer-
tainty
0.00 M ... 50.0 G 0.60 M ...10.0 G 100 V ...250 V ±(10% rdg. + 5 d) ± (15% rdg. + 7 d)
>10.0 G...50.0 G ±(25% rdg. + 5 d) ± (50% rdg. + 20 d)
0.00 M ... 250 G 0.40 M...50.0 G >250 V ... 1.00 kV ±(7% rdg. + 5 d) ± (10% rdg. + 7 d)
>50.0 G...250 G ±(25% rdg. + 5 d) ±(50% rdg. + 20 d)
0.00 M ... 999 G 0.40 M...200 G >1.00 kV ... 5.00 kV ±(7% rdg. + 5 d) ±(10% rdg. + 7 d)
>200 G...999 G ±(25% rdg. + 5 d) ±(50% rdg. + 20 d)
t1 t2 Limit
PI
00:00 ... 01:00 ... 99:50 min 00:00 ... 10:00 ... 99:50 min 0.10 ... 4.00 ... 9.80 min
DAR
00:00 ... 00:30 ... 99:50 min 00:00 ... 01:00 ... 99:50 min 0.10 ... 1.60 ... 9.80 min
Nominal Values,
Test Voltage
Variable
Test Voltage
Nominal Current Intrinsic Uncertainty
100 V, 250 V, 500 V, 1.00 kV
1.0 mA 0 ... +25% rdg.
1.50 kV, 2.00 kV, 2.50 kV 0.4 mA ± 5% rdg.
5.00 kV 0.1 mA ± 3.5% rdg.
100 V...1.00 kV 1.0 mA ± 15% rdg.
> 1.00 kV...2.50 kV 0.4 mA ± 5% rdg.
> 2.50 kV...5.00 kV 0.1 mA ± 3.5% rdg.
Measuring Range Frequency
in Hz
Impedance Intrinsic
Uncertainty
Measuring
Uncertainty
dc test voltage
50 V ... 5.00 kV
±(2.5% rdg. + 5 d) ±(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc 15 ... 500 1 MΩ ±(2.5% rdg. + 2 d) ± (5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc >500 ... 1 k 1 MΩ ±(10% rdg. + 2 d)
±(12.5 % rdg. + 5 d)
Measuring Range Impedance Intrinsic
Uncertainty
Measuring
Uncertainty
15.0 Hz 1.00 kHz 1 MΩ±(0.5% rdg. + 2 d) ±(1% rdg. + 2 d)
Parameter Setting Range Intrinsic Uncertainty Measuring
Uncertainty
Voltage range 100 5,000 V ± (10% rdg. + 8 d)
±
(
15% rdg. + 10 d)
Rise time 5 300 s
Measuring time 1 120 s /
auto / continuous
measurement
——
GMC-I Messtechnik GmbH 55
Capacitance Measurement
Dielectric Discharge (DD)
Reference Conditions
Ambient temperature +23 °C ±2K
Relative humidity 40 ... 60%
Measured quantity
frequency 50 Hz ±10 Hz (during voltage
measurement)
Line voltage
waveshape Sinusoidal, deviation between RMS and
rectified value < 1%
Power Supply MERISO PRIME+
Line voltage 207 V
253 V / 49 Hz 61 Hz
or (depending on country-specific design)
108 V
132 V / 59 Hz 61 Hz
Power consumption < 18 VA
Set of batteries NiMH 9.6 V, 3 Ah, charging time 6 hours
Number of measurements
with nominal current
per VDE 0413 700
Power Supply PROFITEST 204HP/HV
Line voltage 207 V
253 V / 49 Hz 61 Hz
Power consumption PROFITEST 204HP-2.5kV: max. 700 VA
PROFITEST 204HV-5.4kV: max. 100 VA
Electrical Safety
Standards IEC 61010-1:2001
EN 61010-1:2002
VDE regulations VDE 0411 Part 1:2002
Pollution degree 2
Protection IP 40
– MERISO PRIME+
Measuring category
Insulation measurement – 5000 V DC –
no overvoltage
Voltage measurement – 1000 V – CAT II
Safety class II
– PROFITEST 204HP/HV
Safety class I
Safety shutdown if instrument overheats
Fuses mains: F 3.15 / 250
Display
Range
Measuring
Range
Test Voltage Intrinsic Uncertainty Measuring
Uncertainty
0.00
10.0
μ
F 0.10
5.00
μ
F
100450 V ±(10 % rdg. + 5 d) ± (15% rdg. + 8 d)
5005 kV ± (5 % rdg. + 5 d) ± (10% rdg. + 8 d)
Limit
DD 0.10 ... 2.00 ... 9.80
56 GMC-I Messtechnik GmbH
Electromagnetic Compatibility (EMC) MERISO PRIME+
Product standards EN 61326-1:2006
Warning!
This is a class A device. This device may cause radio
interference in residential areas.
If this is the case, the owner may be required to imple-
ment appropriate corrective measures.
Ambient Conditions
Accuracy 0 °C ... + 40 °C
Operating temperature 5 °C ... + 40 °C
Storage temperature 20 °C ... + 60 °C (without battery set)
Relative humidity max. 75%, no condensation allowed
Elevation to 2000 m
Mechanical Design MERISO PRIME+
Display Multiple display with dot matrix
128 x 64 pixels
Dimensions W x H x D:
255 mm x 133 mm x 240 mm
Weight approx. 5 kg with batteries
Mechanical Design
MERISO PRIME+ and PROFITEST 204HP/HV (Feature B1/B2)
Dimensions PROFITEST 204HP/HV:
W x D x H: 254 mm x 130 mm x 285 mm
MERISO PRIME+ and PROFITEST 204HP/HV
completely mounted on Caddy204:
W x D x H: 380 mm x 250 mm x 650 mm
Weight PROFITEST 204HP/HV: approx. 8 kg
MERISO PRIME+ and PROFITEST HP/HV:
completely assembled: approx. 13 kg
Feature B1: PROFITEST 204HP-2.5kV
Feature B2: PROFITEST 204HV-5.4kV
Interference Emission
EN 55022 Class A
Interference Immunity Te st Va lu e
EN 61000-4-2
Contact/atmos. – 4 kV/8 kV
EN 61000-4-3 10 V/m
EN 61000-4-4
Mains connection – 2 kV
EN 61000-4-5
Mains connection – 1 kV
EN 61000-4-6 Mains connection
3 V
EN 61000-4-11 0.5 period / 100%
Nominal Range of
Use
Reso-
lution
Measuring
Uncertainty
Intrinsic Uncertainty
Test vo lt ag e
U AC
250 V ... 2.5 kV 1 V
10 V
±(5% rdg. + 5 d) ±(2.5% rdg. + 5 d)
Measured Quantity
Current I AC
10.0 ... 200 mA 0.1 mA
1mA
±(7% rdg.+ 5 d) ±(5% rdg. + 5 d)
Nominal Range of
Use
Resolu-
tion
Measuring
Uncertainty
Intrinsic Uncertainty
Test vo lt ag e
U AC
650 V ... 1.00 kV
1.00 kV ... 5.35 kV
1V
10 V
+2 ... –7% rdg.
+2 ... –5% rdg.
0 ... –5% rdg.
0 ... –3% rdg.
Measured Identity.
Current I AC
1.0 ... 10.0 mA 0.01 mA
0.1 mA
±(7% rdg.+ 5 d) ±(5% rdg. + 5 d)
GMC-I Messtechnik GmbH 57
8 Data Interfaces
8.1 RS 232 Serial Interface
The data interface socket connector (4) is used to connect the
SECUTEST PSI module (feature I1), a barcode scanner (acces-
sory no. B3261) or the PROFI-MFII interface adapter (accessory
no. Z504H) for entering data with a PC keyboard.
Test results can be printed out at the SECUTEST PSI as a report
by listing measured values only.
An IBM compatible PC can also be connected to this interface,
thus allowing for data exchange.
Pins 9 and 6 may not be short jumpered to pin 5!
8.1.1 Software Analysis of Measurement Results
Test reports can be conveniently created at a PC and down-
loaded to the test instrument with the included software.
8.1.2 Interface Specification and Protocol
The interface included with the MERISO PRIME+ complies with
the RS 232 standard.
Technical Data:
Baud rate 9600 baud, fixed
Character length 8 bit
Parity none
Stop bits 1
All MERISO PRIME+ control functions can also be remote con-
trolled via this interface. Signal and display values can be read out
as well.
A description of the interface protocol, as well as commands and
syntax, can be requested from GMC-I Messtechnik GmbH.
8.1.3 Connector Pin Assignments
1: NC
2: TxD (PSI)
3: RxD (PSI)
4: NC 6
5: GND
6: NC 4
7: NC 8
8: NC 7
9: +9 V
9876
5 4321
RxD TxD
GND
+9V +5V
58 GMC-I Messtechnik GmbH
9 Displays and Error Messages – Causes – Remedies
Displays and Error Messages Meaning / Cause Remedy
MERISO PRIME+
Test Sequence
Countdown runs here until residual voltage is less than 60 V.
All-clear after an insulation test:
voltage at test probes less than 25 or 60 V depending upon measuring
function
The test has been passed.
The test results are not within allowable limit values.
LCD does not light up
– Mains cable not plugged in
– Mains fuse is blown
– Plug in the mains cable (see chapter 3.4 on page 20).
– Replace the mains fuse (see chapter 10.1 on page 61).
Indicates type of voltage:
Alternating voltage with superimposed direct voltage – direct voltage –
alternating voltage
Insulation measurement:
Interference voltage at the test probes, measurement aborted
Warning against high voltage,
voltage at the test probe between 1,000 V and 5,000 V
Warning concerning a source of danger,
voltage at the test probe between 25 V and 1,000 V
This symbol appears after the voltage measurement value has been fro-
zen (see chapter 4.6 on page 46).
Database Operations
– No measurement has been performed.
– The measurement has already been saved to memory.
See text at the display. Press the START key to repair.
GMC-I Messtechnik GmbH 59
An error has occurred while saving data to memory. Save data to a PC and then clear the memory.
See text at the display. Press the START key to repair.
Memory in no longer able to save new measurement data. Save data to a PC and then clear the memory.
See text at the display. Transmit data to a PC or print them out, and then delete
DUTs which are no longer necessary, or clear the entire
memory.
No data have yet been saved to this data record.
Printing is not possible. Check connections at the printer and make sure paper is
available.
Check the printer for possible errors.
Data are being transferred to the printer. The printer may wait until data
transmission has been entirely completed before it starts printing.
A filled file cabinet indicates that measurement data have already been
saved to memory for the selected DUT.
The data record has been deleted for a DUT.
The printer is not working. Check the data record.
The printer is not working. Check the connection between the printer and the MERISO
PRIME+.
Then press the START key.
The printer is working correctly. Turn the selector switch in order to abort the print-out.
Displays and Error Messages Meaning / Cause Remedy
60 GMC-I Messtechnik GmbH
The measurement has been saved to memory.
MERISO PRIME+(with feature B1 or B2)
The displayed high-voltage module has been detected. Turn device on with key switch.
The displayed high-voltage module has been detected. Turn device on with key switch.
Green signal lamps do not light up.
The displayed high-voltage module has not been detected.
– Mains cable not plugged in
– Key switch set to “0”
– Blown fuse
– Defective signal lamp
– Plug in the mains cable.
– Set the key switch to “I”.
– Replace the fuse
– Service
Red signal lamps do not light up.
– No data transfer between MERISO PRIME+ and 204HP/HV
– Defective signal lamp
– See “Device defective!” error message.
– Replace signal lamp, service
High-voltage module ready to be switched on
High-voltage is present at the test probes during measurement.
Do not touch the test probes!
The specified limit value I
MAX
has been exceeded. Current limiting has
switched the device to the “standby” mode.
Test completed successfully. Residual voltage less than 25 V
Phase angle display
Continuous operation with high test power during high voltage testing Allow the device to cool for approximately 10 minutes.
One of the two signal lamps in the high-voltage module is defective. Send the high-voltage module to GMC-I Service GmbH for
repair.
Displays and Error Messages Meaning / Cause Remedy
or
GMC-I Messtechnik GmbH 61
10 Maintenance
10.1 Replacing the Mains Fuse
Attention!
!
Disconnect the device from the measuring circuit at all
poles before opening the fuse compartment!
Disconnect the device from the mains.
Attention!
!
Use only specified, original replacement fuses!
Bypassing or repairing fuses is prohibited!
If fuses with other current ratings, breaking capacities or
blowing characteristics are used, the device may be
damaged!
Ð Open the fuse compartment at the inlet socket with a suitable
tool (e.g. a screwdriver).
Ð Replace the blown fuse with a new fuse with identical specifi-
cations.
Ð Close the fuse compartment.
10.2 Testing of Rechargeable Batteries
Check the batteries for leakage or damage at short regular inter-
vals, and after long periods of storage. If damage or leakage has
occurred, carefully and completely clean the electrolyte from the
instrument with a damp cloth, and replace the batteries with a
new battery pack (accessory no. …..) before placing the instru-
ment back into service.
If the battery test (see chapter 3.7.4 on page 25) indicates that backup
battery voltage or rechargeable battery voltage has dropped below the
minimum allowable value, recharge the battery pack or have the backup
battery replaced by GMC-I Service GmbH.
Note
Prior to lengthy periods of rest (e. g. holiday), we recom-
mend removing the (rechargeable) batteries. This helps to
prevent excessive depletion or leakage of batteries,
which, under unfavourable circumstances, may cause
damage to the instrument.
10.3 Housing and Test Probes
No special maintenance is required. Keep outside surfaces clean
and dry. Use a slightly dampened cloth for cleaning. Avoid the use
of solvents, cleansers and abrasives.
Attention!
!
No condensation may be permitted to occur at the test
instrument, the test cables or the device under test,
because high-voltage may otherwise cause leakage cur-
rent at the surfaces. Insulated components may also
conduct high-voltage in this case.
Measurements may not be performed under moist ambient con-
ditions, where condensation has occurred or in explosive atmo-
spheres.
62 GMC-I Messtechnik GmbH
10.4 Measurement Cables
Inspect the measurement cables for mechanical damage at regu-
lar intervals.
Attention!
!
Even if only minimal damage is discovered at the test ca-
bles, it is advisable to send them immediately to GMC-I
Service GmbH.
10.5 High-Voltage Module Test Cables (Feature B1/B2)
Wind the test cables onto the provided cable uptakes at the back
of the trolley (feature D1) after each use if included. Test cables
may not, under any circumstances, be mechanically damaged or
kinked, because impaired insulation characteristics may other-
wise result.
Inspect the test cables and the high-voltage pistols for mechani-
cal damage before placing the high-voltage module into service
each time it is used.
Attention!
!
Even if only minimal damage is discovered at the test ca-
bles or high-voltage pistols, it is advisable to send them
immediately to the GMC-I Messtechnik GmbH Repair
and Replacement Parts service department.
10.6 Replacing Bulbs in the External Signal Lamps (Feature F1)
Attention!
!
Before replacing bulbs in the external signal lamps, dis-
connect them from the connector terminals at the test
instrument.
Ð Unscrew the red or green lens by turning it counterclockwise.
Ð Remove the defective bulb from the bayonet socket and insert
a new 12 to 15 V / 2 W bulb with BA9s base, e.g. OSRAM
miniwatt T10 no. 3453B.
Ð Screw the lens back into place by turning it clockwise.
10.7 Software
Internal test instrument software can be updated with the help of
a PC and an interface cable for the RS 232 interface. The soft-
ware is downloaded to the flash EPROM in the MERISO
PRIME+.
The data file with the desired software version is transmitted to
the test instrument via the serial interface with the help of the
update function included with WinProfi software. Previously
installed software is overwritten in the process.
Downloading Procedure
Ð Establish a connection between the PC and the
MERISO PRIME+.
Ð Switch the PC and the test instrument on.
Continue as described in chapter 3.7.5 on page 26.
GMC-I Messtechnik GmbH 63
Note
After updating the software, reinstall the desired lan-
guage to the test instrument as well, even if the language
will not be changed. This assures that all of the additional
functions included in the update are correctly displayed.
10.8 Recalibration
The respective measuring task and the stress to which your mea-
suring instrument is subjected affect the ageing of the compo-
nents and may result in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is fre-
quently used in field applications, combined with transport stress
and great temperature fluctuations, we recommend a relatively
short calibration interval of 1 year. If your measuring instrument is
mainly used in the laboratory and indoors without being exposed
to any major climatic or mechanical stress, a calibration interval of
2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in rela-
tion to traceable standards are measured and documented. The
deviations determined in the process are used for correction of
the readings during subsequent application.
We are pleased to perform DAkkS or factory calibrations for you in
our calibration laboratory. Please visit our website at
www.gossenmetrawatt.com ( Company DAkkS Calibration
Center or FAQs Calibration questions and answers).
By having your measuring instrument calibrated regularly, you ful-
fill the requirements of a quality management system per
DIN EN ISO 9001.
* Verification of specifications or adjustment services are not part of the calibration.
For products from our factory, however, any necessary adjustment is frequently
performed and the observance of the relevant specification is confirmed.
10.9 Device Return and Environmentally Compatible Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German Electrical and
Electronic Device Law). This device is subject to the RoHS direc-
tive.
We identify our electrical and electronic devices in accor-
dance with WEEE 2012/19EU and ElektroG with the sym-
bol shown to the right per DIN EN 50419.
These devices may not be disposed of with the trash. Please con-
tact our service department regarding the return of old devices.
If you use batteries or rechargeable batteries in your instrument or
accessories which no longer function properly, they must be duly
disposed of in compliance with the applicable national regula-
tions.
Batteries or rechargeable batteries may contain harmful sub-
stances or heavy metal such as lead (PB), cadmium (CD) or mer-
cury (Hg).
They symbol shown to the right indicates that batteries or
rechargeable batteries may not be disposed of with the
trash, but must be delivered to collection points specially
provided for this purpose.
Pb Cd Hg
64 GMC-I Messtechnik GmbH
11 Appendix
11.1 Glossary
11.2 Minimum Display Values in Consideration of
Measuring Error
Table for determining minimal display values for insulation resis-
tance in consideration of the instrument’s measuring error:
Abbreviation
Meaning
DAR
Absorption Ratio: Ratio between insulation resistance values measured after 30 and 60 seconds
DD Dielectric Discharge
ΔR
Lim
Maximum allowable statistical deviation from the measured mean value
Δt/250V Dwell Time per Step
PI Polarization Index:
Ratio between insulation resistance values measured after 1 and 10 minutes
I
MAX
Max. allowable current before high-voltage is switched off
I
T
Interrupting current for high-voltage test
R
ins
Insulation Resistance
Limit R
ins
Limit value for insulation resistance
T
R
Rise time: time during which test voltage climbs to U
AC.MAX.
T
D
Test duration at maximum test voltage U
AC.MAX.
(not including rise time TR)
U
AC.MAX.
Maximum test voltage for high-voltage testing
U
brk
Breakdown Voltage
U
ins
Test Vo lt ag e
U
(ins) end
Ramp Function Upper Limit
U
meas
Measuring Voltage
U
nom
Nominal Test Voltage
U
T
Voltage value before exceeding interrupting current during high-voltage testing
U
STEP
Test voltage as ramp sequence
U
test
Test Vo lt ag e
U
var
Variable Test Voltage
Limit Value Minimum Display
Value
Limit Value Minimum Display
Value
0.10 MΩ 0.12 MΩ 100 MΩ 118 MΩ
0.20 MΩ 0.23 MΩ 200 MΩ 228 MΩ
0.50 MΩ 0.56 MΩ 500 MΩ 558 MΩ
1.00 MΩ 1.18 MΩ 1.00 GΩ 1.18 GΩ
2.00 MΩ 2.28 MΩ 2.00 GΩ 2.28 GΩ
5.00 MΩ 5.58 MΩ 5.00 GΩ 5.58 GΩ
10.0 MΩ 11.8 MΩ 10.0 GΩ 11.8 GΩ
20.0 MΩ 22.8 MΩ
50.0 MΩ 55.8 MΩ
GMC-I Messtechnik GmbH 65
11.3 Voltage at Device Under Test During Insulation Resistance Test
Test voltage U at the device under test depending upon its
resistance R
x
at nominal voltage 100, 500, 1000, 2400 and
5000 V:
100
U/V
R
x
/kΩ
100
80
60
40
20
0
U
N
= 100 V
0
100
200
300
400
500
500
U
N
= 500 V
U/V
R
x
/kΩ
0
200
400
600
800
1000
1
U
N
= 1000 V
U/V
R
x
/MΩ
0
500
1000
1500
2000
2400
3
U
N
= 2400 V
U/V
R
x
/MΩ
1
0
1000
2000
3000
4000
5000
5
U
N
= 5000 V
U/V
R
x
/MΩ
10 2031
2400
3200
66 GMC-I Messtechnik GmbH
11.4 Index
B
Beeper
acoustic . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
C
Current limiting in the event of arc-over . . . . 13
D
Data management
Data
Deleting . . . . . . . . . . . . . . . . . . . . . . . 51
Processing . . . . . . . . . . . . . . . . . . . . 50
Description
Copying . . . . . . . . . . . . . . . . . . . . . . . 50
Deleting . . . . . . . . . . . . . . . . . . . . . . . 50
Entry . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Entry options . . . . . . . . . . . . . . . . . . 49
Display messages . . . . . . . . . . . . . . . . . . . 58
Memory
Deleting . . . . . . . . . . . . . . . . . . . . . . . 52
Occupancy . . . . . . . . . . . . . . . . . . . . 48
Testing . . . . . . . . . . . . . . . . . . . . . . . . 51
Printing test results . . . . . . . . . . . . . . . . . . 53
Selecting device under test . . . . . . . . . . . 48
Discharge
Automatic after insulation test . . . . . . . . 29
Dielectric during capacitance measurement
.44
E
Electrical isolation . . . . . . . . . . . . . . . . . . . . . . . 13
F
Fuses
Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Replacement . . . . . . . . . . . . . . . . . . . . . . . 61
I
Interference voltage
during insulation test . . . . . . . . . . . . . . . . . 29
Interrupting current
Display messages . . . . . . . . . . . . . . . . . . . 13
in case of breakdown voltage . . . . . . . . . 38
in pulse control mode . . . . . . . . . . . . . . . . 39
M
Measuring of
Absorption index . . . . . . . . . . . . . . . . . . . . 40
Breakdown voltage . . . . . . . . . . . . . . . . . . 42
Capacitance . . . . . . . . . . . . . . . . . . . . . . . . 44
High resistances . . . . . . . . . . . . . . . . . . . . 33
Polarization index . . . . . . . . . . . . . . . . . . . 40
Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
P
Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Product Support Hotline . . . . . . . . . . . . . . . . . 68
R
Residual voltage
Determining the value . . . . . . . . . . . . . . . . 46
Display of discharge time . . . . . . . . . . . . . 46
S
Safety measures
Control circuits . . . . . . . . . . . . . . . . . . . . . . 11
Frequency Converters . . . . . . . . . . . . . . . . 11
Neutral conductor . . . . . . . . . . . . . . . . . . . 11
PELV circuits . . . . . . . . . . . . . . . . . . . . . . . . 11
Persons . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
TN systems . . . . . . . . . . . . . . . . . . . . . . . . . 11
Signal lamp
optical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Statistical analysis
during insulation measurement . . . . . . . . 32
Storage batteries
Display of measuring voltage . . . . . . . . . . 25
Display of remaining capacity . . . . . . . . . 20
Operation with battery pack . . . . . . . . . . 15
Recharging during mains operation . . . . 20
T
Test voltage
during HV test
in pulse control mode . . . . . . . . . . . 39
on device under test . . . . . . . . . . . . 65
during insulation test
Fixed steps . . . . . . . . . . . . . . . . . . . . 31
on device under test . . . . . . . . . . . . 65
variable/50 V steps . . . . . . . . . . . . . . 31
during insulation test as a sequence of
steps . . . . . . . . . . . . . . . . . . . . . 31
W
WinProfi
Installation and program start . . . . . . . . . 26
Transmission and print-out of report data
. . . 28
GMC-I Messtechnik GmbH 67
12 Repair and Replacement Parts Service,
Calibration Center* and Rental Instrument Service
When you need service, please contact:
GMC-I Service GmbH
Service Center
Thomas-Mann-Str. 20
90471 Nürnberg, Germany
Phone: +49 911 817718-0
Fax: +49 911 817718-253
e-mail: service@gossenmetrawatt.com
www.gmci-service.com
This address is only valid in Germany.
Please contact our representatives or subsidiaries for service in
other countries.
* DAkkS Calibration Laboratory for Measured Electrical Quantities:
D-K-15080-01-01 accredited as per DIN EN ISO/IEC 17025:2005
Accredited quantities: direct voltage, direct current value, direct current resistance,
alternating voltage, alternating current value, alternating current active power,
alternating current apparent power, DC power, capacitance, frequency and
temperature
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with
DIN EN ISO 9001:2008.
Our DAkkS calibration laboratory is accredited by the Deutsche
Akkreditierungsstelle GmbH (National accreditation body for the
Federal Republic of Germany) in accordance with DIN EN ISO/IEC
17025:2005 under registration number D-K-15080-01-01.
We offer a complete range of expertise in the field of metrology:
from test reports and proprietary calibration certificates right on up to
DAkkS calibration certificates.
Our spectrum of offerings is rounded out with free test equipment
management.
An on-site DAkkS calibration station is an integral part of our service
department. If errors are discovered during calibration, our spe-
cialized personnel are capable of completing repairs using original
replacement parts.
As a full service calibration laboratory, we can calibrate instru-
ments from other manufacturers as well.
Edited in Germany Subject to change without notice A pdf version is available on the internet.
GMC-I Messtechnik GmbH
Südwestpark 15
90449 Nürnberg •
Germany
Phone +49 911 8602-111
Fax +49 911 8602-777
E-Mail info@gossenmetrawatt.com
www.gossenmetrawatt.com
13 Product Support
When you need service, please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone: +49 911 8602-0
Fax: +49 911 8602-709
e-mail support@gossenmetrawatt.com
66


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME plus in the language / languages: English as an attachment in your email.

The manual is 1,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME plus

Gossen Metrawatt METRISO PRIME plus User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info