10 GMC-I Messtechnik GmbH
2 Safety Features and Precautions
The MERISO PRIME+ test instrument, and the
PROFITEST204HP/2.5kV and PROFITEST204HV/5.4kV high
voltage modules have been manufactured and tested in accor-
dance with the following safety regulations:
Standard DIN EN 61557-1:2007
DIN EN 61557-2:1998-05
IEC 61010-1:2004
EN 61010-1:2002
VDE regulation VDE 0413 Part 1:2007
VDE 0413 Part 2:1998-05
VDE 0411 Part 1:2002
If used for its intended purpose, safety of the user, the instrument
and the device under test is assured.
Read the operating instructions completely and carefully before using the
device, and follow all instructions included therein.
The operating instructions should be made available to all users.
Observe the following safety precautions:
The instrument may only be connected to a 230 V or 120 V
electrical system (depending on the country-specific design)
which is protected with a fuse or circuit breaker with a maxi-
mum nominal current rating of 16 A.
Be prepared for the occurrence of unexpected voltages at de-
vices under test (for example, capacitors may be dangerously
charged).
Make sure that connector, measurement and test cables are
not damaged (i.e. broken insulation, kinks, interruptions etc.).
The instrument may only be used indoors.
Impulse withstand voltage of up to 2.5 kV may occur in
measurement category II systems with 230/400 V.
If only the batteries are to be charged (instrument is ready for
operation in SETUP position), please ensure that the test in-
strument cannot be used by unauthorized personnel during
the charging process.
Measurement may only be performed on voltage-free equip-
ment and devices.
Measurements may not be performed under moist ambient
conditions, where condensation has occurred, or in explosive
atmospheres.
Do not disconnect the measurement cables until the device
under test has been fully discharged.
Observe accident prevention regulation BGV A3, “Electrical
Systems and Equipment”.
Do not perform measurements alone: always have a second
person help if possible.
Make sure that measurement cables are connected correctly.
Do not touch the test probes at any point beyond the finger
guard.
The MERISO PRIME+ may not be used:
If visible damage is apparent
With damaged connector, measuring or test cables (test ca-
bles may not, under any circumstances, be mechanically dam-
aged or kinked, because impaired insulation characteristics
may otherwise result)
If these no longer function flawlessly
If the signal lamps on the PROFITEST204HP/2.5kV
(feature B1) and PROFITEST204HV/5.4kV (feature B2) high-
voltage modules are defective
In such cases, the devices must be removed from service and
secured against any possible inadvertent use.
Send the instrument to our Repair and Replacement Parts service
department in such cases (see chapter 12 on page 67).
10


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO PRIME plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO PRIME plus in the language / languages: English as an attachment in your email.

The manual is 1,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRISO PRIME plus

Gossen Metrawatt METRISO PRIME plus User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info