Ð Turn the device over so that the control panel faces up and
remove the housing top (i.e. control panel).
Ð Remove the fuse from its holder with the help of an object such
as a test probe, and replace it with a new fuse.
(A replacement fuse is available in the carrying pouch).
Note!
Replace the fuse only in a clean and dust-free environment.
Soiling may possibly impair the correct functioning of the
measuring instrument.
Attention!
!
Use specified fuses only! If fuses with other blowing
characteristics, other current ratings or other breaking
capacities are used, the operator is placed in danger, and
resistors and other components may be damaged.
The use of repaired fuses or short-circuiting the fuse holder
is prohibited.
Ð Set the housing top back into place without applying pressure.
Ð Lay the carrying strap over the metal pin.
Note!
Make sure that the cable from the measuring element is not
pinched when the housing bottom and the housing top are
fitted together.
Ð Position the measuring instrument so that you view it from the
side with the test probe snapped in to the detaining fork. The
support included on the housing top must slide into the guide in
proximity to the test probe holder, and may not be tilted at the
display module. In this way, the housing top can be pressed
onto the housing bottom without applying excessive force.
Ð Hold the housing top and bottom together and turn the entire
instrument over so that the housing bottom faces up.
Ð Retighten the screws.
5.4 Recalibration
The respective measuring task and the stress to which your
measuring instrument is subjected affect the ageing of the
components and may result in deviations from the guaranteed
accuracy.
If high measuring accuracy is required and the instrument is
frequently used in field applications, combined with transport stress
and great temperature fluctuations, we recommend a relatively short
calibration interval of 1 year. If your measuring instrument is mainly
used in the laboratory and indoors without being exposed to any
major climatic or mechanical stress, a calibration interval of 2-3 years
is usually sufficient.
During recalibration
*
in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in relation
to traceable standards are measured and documented. The
deviations determined in the process are used for correction of the
readings during subsequent application.
*
Verification of specifications or adjustment services are not part of the
calibration. For products from our factory, however, any necessary
adjustment is frequently performed and the observance of the relevant
specification is confirmed.
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO 5024 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO 5024 in the language / languages: English as an attachment in your email.

The manual is 0,17 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info