5 Maintenance – Recalibration
5.1 Housing
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the
use of cleansers, abrasives and solvents.
5.2 Batteries
Check the batteries at short regular intervals to makes sure no
leakage has occurred. If leakage occurs, electrolyte must be fully
removed from the instrument and new batteries must be installed.
Remove the batteries from the battery compartment if the measuring
instrument will not be used for a lengthy period of time. Refer to
chapter 3.2 regarding battery replacement
Attention: If the LED lights up red before or during measurement, the
batteries must be replaced immediately.
Use only recommended battery types (see chapter 3.2).
5.2.1 Details on Power Consumption and Operating Time
Low-Resistance Measurement
The instrument draws the most power from the batteries when the
value of the contact resistance is the lowest.
Power consumption does not exceed the value of 0.4 A.
Insulation Resistance Measurement
Power consumption from the batteries depends on the selected test
voltage and the actual value of the insulation resistance, i.e. on the
load imposed on the source of the measuring voltage U
N.
The table
below specifies the power consumption from the batteries for the
indicated parameters.
The following table lists the number of possible measurements with
one set of alcaline batteries by taking into account the power con-
sumption from the batteries.
The number of possible measurements depends on the following conditions:
one measurement takes 1 minute
the insulation resistance complies with the requirements of the
standard
(power consumption from the battery in accordance with
column 3 of the table above).
5.3 Replacing the Fuse
The low-resistance measuring range is protected with a fuse, see
chapter 6. The fuse blows if a voltage is applied to the test probes
which may cause damage to the instrument as a result of its
magnitude or polarity.
Adhere to the prescribed sequence when replacing the fuse:
Ð Remove the test probes from the device under test.
Ð Loosen the 3 screws in the housing base.
Test Voltage
U
N
R
X
measured at I
BAT
(U
BAT
= 6 V)
(*if LED lights up green, pointer oscillates and there is no acoustic signal,
the transformer is overloaded)
500 V 0 kΩ / < 1 A * 0.1 MΩ / < 1.1 A > 1 MΩ / < 0.6 A
> 400 M
Ω
/ < 0.3 A
250 V 0 kΩ / < 1 A * 0.1 MΩ / < 0.4 A
> 500 k
Ω
/ < 0.4 A > 400 M
Ω
/ < 0.2 A
100 V 0 kΩ / < 1 A * 0.1 MΩ / < 0.2 A
> 200 k
Ω
/ < 0.3 A > 400 M
Ω
/ < 0.2 A
Test Voltage U
N
AlMn Battery min. 2300 mAh
500 V > 300 measurements
250 V > 500 measurements
100 V > 750 measurements
10


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRISO 5024 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRISO 5024 in the language / languages: English as an attachment in your email.

The manual is 0,17 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info