14 GMC-I Messtechnik GMBH
automatically adjusted, the measuring ranges should not
be changed manually.
The DATA function is deactivated if the DATA | CLEAR key is
pressed and held (approx. 1 s) or by switching the instru-
ment off and back on again.
7 Milliohm Measurement (4-Pole-Measurement)
7.1 Compensation of Cable Resistance
Electrical resistance is a dipole quantity which can
generally only be measured using two poles. This is
accomplished by directing a measuring current of
predetermined magnitude through the device under test,
and measuring the resultant voltage drop. The respective
resistance value is derived from the quotient of these two
values.
The two points between which voltage is measured are
decisive as regards the results of the measurement. All
resistances between these two points add to the
measured resistance value. These include contact
resistance, as well as cable resistance. If a very low
resistance value needs to be measured, for example
contact resistance at a contactor with a value of only a few
milliohms, the points between which voltage is measured
must be moved out of the measuring instrument and
positioned as closely as possible to the device under test.
For this reason, the measuring instrument is equipped with
separate jacks for current feed and voltage measurement.
This type of 4-pole connection is known as connection
according to Kelvin.
KC4 Kelvin clips and KC27 Kelvin probes (available as
accessories) allow for easy, correct connection.
7.1.1 Measurement with Kelvin Probe KC27
m
U+U–
I+
red
black
black
Milliohm measurement on
conductor bars, welds,
coils, etc. with Kelvin probe
KC27 and Kelvin clip KC4
Direct the spring-loaded test
probe inward towards the DUT,
if possible, as it is connected
via the red plug with U
+
or U
(Sense) .
I–
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT 27EX at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT 27EX in the language / languages: English as an attachment in your email.

The manual is 0,4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info