Adjusting Cable Resistance
Ð After selecting a resistance thermometer, the
measurement display is accessed by pressing the key.
Ð The menu for cable resistance adjustment is opened by
simultaneously pressing the and keys. Rlead and
temp are displayed.
Ð The decade (i.e. the position of the digit to be changed) is
selected with the keys, and the respective digit is
set with the keys.
Ð The menu is exited upon acknowledging the last digit by
pressing the key, whereupon the measuring display
returns. The cable resistance value remains in memory.
The default value is 0.1 . Entry is limited to a range of 0
to 50 .
Note!
The selected cable resistance value remains
unchanged after exiting the function, and when the
instrument is switched off.
11 Insulation Resistance Measurement [M@...V]
(METRA HIT 27I and METRA HIT H+E
CAR only)
11.1 Preparing for Measurement
Note!
High-resistance measurements
In the event of extremely high resistances the
capacitive influence of the person performing the
measurement and/or the measuring cable may
distort the measured value. Use therefore short or
shielded measurement cables.
The measurement cables may not come into contact
with one another during high-resistance insulation
measurements.
Ð Set the rotary selector switch to “M@50V, “100V”,
“250V” or “500V”, depending upon the desired test
voltage.
Ð Connect the DUT as shown on the next page.
Note!
Interference voltage
The rotary selector switch may only be set to
M@...V for insulation resistance measurement
(not for voltage measurement).
However, if interference voltage is inadvertently
applied with the switch in this position, it is displayed
at the bottom left. Insulation resistances may only be
measured on voltage-free objects.
If an interference voltage of 50 V is present, insulation
resistance measurement is disabled. The LCD continues to
display the interference voltage value. If a voltage of greater
than 610 V is present, an acoustic signal is generated as well.
m
S+S–
VC
RTD
see chapter 14
Measuring Range:
- - - - -
V
-000 250 V
M@
50 ... 500V
10.29
V
023 250 V
M
Selected Test Voltage
Displays test voltage
as long as key is depressed
R
ISO
FUNC MAN
+
FUNC
Continuous
Measurement
Measuring voltage
/Residual voltage
approx. 2% accuracy
Nominal/
test voltage
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRA HIT H-E CAR at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRA HIT H-E CAR in the language / languages: English as an attachment in your email.

The manual is 0,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRA HIT H-E CAR

Gossen Metrawatt METRA HIT H-E CAR Quick start guide - English, German, French - 4 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info