590606
12
Zoom out
Zoom in
Previous page
1/28
Next page
12 GMC-I Messtechnik GmbH
8.2 Continuity Test during Resistance Measurement
If the “acoustic signal ” function is activated and the
0 ... 310 Ω measuring range is selected, a continuous
acoustic signal is generated by the instrument within a range
of 0 to approx. 10 Ω.
Overflow “0l is displayed where Rd > 310 Ω.
Activating and Deactivating Continuity Testing (acoustic signal)
Ð Set the rotary selector switch to „Ω//.
Ð Repeatedly press the FUNC key until and “short” appear
at the display, provided the connector jacks are open
(Display 0.L).
Ð Connect the measurement cables to the device under
test.
Ð The instrument is switched to resistance measurement
by pressing the FUNC key once again.
8.3 Diode Measurements
Ð Make sure that the device under test is voltage-free (see
chapter 7.1). Interference voltages distort measurement
results!
Ð Set the rotary selector switch to „Ω//.
Ð Briefly press the FUNC key for diode measurement. Unit of
measure “V” and “diode” are then displayed if the
connector jacks are open (Display 0.L).
Ð Connect the device under test as shown.
Conducting Direction and Short-Circuit
The instrument displays conducting-state voltage in volts. As
long as voltage drop does not exceed the maximum display
value of 3 V, several series connected components or
reference diodes can be tested.
Reverse Direction or Interruption
Overload (.0l) is displayed at the instrument for diode
measurements where Ud > 3.1 V. Measuring current is
always a constant current of approximately 1 mA.
Note!
Resistors and semiconductors which are connected
in parallel to the diode distort measurement results!
mΩ
S+S–
VΩ°C
Conducting Direction
Reverse Direction
12


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRA HIT 27I at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRA HIT 27I in the language / languages: English as an attachment in your email.

The manual is 0,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRA HIT 27I

Gossen Metrawatt METRA HIT 27I Quick start guide - English, German - 4 pages

Gossen Metrawatt METRA HIT 27I User Manual - Dutch - 28 pages

Gossen Metrawatt METRA HIT 27I Quick start guide - French - 4 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info