776255
6
Zoom out
Zoom in
Previous page
1/12
Next page
6GMC-I Messtechnik GmbH
and safety class I devices with accessible
conductive parts which are not connected to the
protective conductor. Contact all accessible
conductive parts at the device under test with the
test probe at the end of the probe cable form the
SECUTEST.../SECULIFE ST to this end.
These measurements must be performed with the plug
in both positions for DUTs with earthing contact plugs!
Note!
This test must be performed in accordance
with the specified test sequence.
Mains power
must be indicated by signal lamps L1, L2 and L3 (9)
before switching the device under test on.
The test outlets (1 – 4) are disconnected
from the mains by the safety shutdown
function included with the AT3-III E for
devices under test with fault currents of
greater than 18 mA during measurement of
protective conductor current with residual
current. Mains power may be disconnected
if the AT3-III E is used in systems protected
with RCCBs rated less than 30 mA. Mains
signal lamps L1, L2 and L3 (9) go out. The
function LED (10) blinks.
Disconnect the AT3-III E from the mains to reset.
Reconnect the AT3-III E to the mains. The test adapter
is once again ready for use after a brief self-test.
The short-circuit test displayed at the
SECUTEST.../SECULIFE ST cannot be performed
due to safety impedances which have been
integrated into the AT3-III E for testing 3-phase
devices.
Figure 1 Connecting Single and 3-Phase Devices to the AT3-III E and the SECUTEST...
230/400 V 50 Hz
test adapter AT3-III E
L3
L2
L1
!
SECUTEST
13
230/400 V 50 Hz
test adapter AT3-III E
L3
L2
L1
!
SECUTEST
13
SECUTEST PRO
COM V 1
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt AT3-III E at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt AT3-III E in the language / languages: English as an attachment in your email.

The manual is 0.59 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info