10 GMC-I Messtechnik GmbH
7Self-Test
Because it is so easy to perform, the self-test should
be conducted before each use of the AT3-III E.
Please be careful to ensure that the case of the
AT3-III E is not being contacted with devices
under PE or earth potential during the self-test.
Performing the Protective Conductor Test
Ð
Insert the mains plug from the
SECUTEST.../
SECULIFE ST
into the earthing contact outlet (12)
at the AT3-III E.
ÐThe earthing contact plug (13) and the SE-
CUTEST port (14) of the AT3-III E may not be
plugged into the SECUTEST.../SECULIFE ST.
ÐConnect the AT3-III E to mains power.
ÐSet the SECUTEST.../SECULIFE ST to the
protective conductor test individual measure-
ment (RSL/RPE).
ÐContact protective conductor potential at the
mains system which is supplying power to the
AT3-III E with the test probe at the end of the
probe cable from the SECUTEST.../
SECULIFE ST, e.g. at the earthing contact of
an earthing contact outlet. If an excessively
high value or interruption is indicated, the pro-
tective conductor is interrupted.
The AT3-III E test adapter must be disconnected
from the mains, and the error in the electrical
system or the test adapter must be corrected. The
AT3-III E is equipped with a “Trigger IΔn” test key in
order to allow for testing the included IΔn safety
shutdown function at any time.
Performing the IΔn Self-Test
ÐRemove the device under test
(device or extension cable).
ÐSet the SECUTEST.../SECULIFE ST to individ-
ual measurement.
ÐSECUTEST.../SECULIFE ST (as of version
3.1): Select the “DI current” sub-menu (resid-
ual current).
ÐSignal lamps L1, L2 and L3 indicate the pres-
ence of line voltage.
ÐActivate the “Trigger IΔn” key (11).
ÐThe AT3-III E disconnects mains power from
the outlets (1 – 4).
ÐSignal lamps L1, L2 and L3 go out.
ÐThe function LED (10) blinks.
Disconnect the AT3-III E from the mains to re-
set. Reconnect the AT3-III E to the mains.
The test adapter is once again ready for use
after a brief self-test.
Note!
If mains power is not switched during this
test, the fuses in the AT3-III E may have
blown. If the AT3-III E still does not function
correctly after replacing the fuses, it must be
removed from service and repaired.
Measurement is no longer possible!
8 Technical Data
Residual Current Measuring Function
Measuring Range 0 20 mA
Transformation Ratio 1 V per 10 mA
Intrinsic Uncertainty ±(5%
of rdg.
+ 0.05 mA)
Nominal Ranges of Use
Line Voltage
L1/L2/L3/N 207 253 V AC
Frequency 49 51 Hz
Temperature 0 °C +40 °C
Line Voltage
Waveshape sinusoidal
Reference Conditions
Ambient Temperature +23 °C ±2K
Relative Humidity 50% ±5%
Line Voltage 230 V/400 V ±10%
Frequency
of Measured Quantity 50 Hz ±0.2%
Ambient Conditions
Operating Temperatures 10 + 40 °C
Storage Temperatures 25 + 60 °C
Humidity max. 75%, no
condensation allowed
Elevation up to 2000 m
Power Supply
Nominal Line Voltage
3~230/400 V/50 Hz/CAT II
Connection only permissible with overload
protection device In = 16 A I2 1.45 In
Electrical Safety
Standard
DIN VDE 0404-1/-2:2002-5
IEC/EN 61010-1:2010
Contamination degree 2
Safety Class I
Residual Current
Shutdown (4-pole) at approx. 18 mA
Device Fuses F0315 L250V
5 x 20 T32mA L 250 V
DIN EN 60127-2
10


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt AT3-III E at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt AT3-III E in the language / languages: English as an attachment in your email.

The manual is 0.59 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email