When the device is in storing , it can’t be turned off.
6 MAINTAIN, TRANSPORT AND STORAGE
6.1 Cleaning and disinfection
The device must be turned off before cleaning, and it should not be immersed into liquid.
Please take out the internal battery before cleaning, do not immerse it into liquid.
Use 75% alcohol to wipe the device enclosure, and use liquid soap or isopropanol to wipe the watch-
band for disinfection, nature dry or clean it with clean and soft cloth. Do not spray any liquid on the
device directly, and avoid liquid penetrating into the device.
6.2 Maintenance
A. Check the main unit and all accessories periodically to make sure that there is no visible damage that
may affect patient’s safety and monitoring performance. It is recommended that the device should be
inspected weekly at least. When there is obvious damage, stop using it.
B. Please clean and disinfect the device before/after using it according to the User Manual (6.1).
C. Please replace the batteries in time when low-battery appears.
D. Please take out the batteries if the device is not used for a long time.
E. The device need not to be calibrated during maintenance.
6.3 Transport and Storage
A. The packed device can be transported by ordinary conveyance or according to transport contract.
During transportation, avoid strong shock, vibration and splashing with rain or snow, and it can not
be transported mixed with toxic, harmful, corrosive material.
B. The packed device should be stored in room with no corrosive gases and good ventilation. Temper-
ature: -40°C~+60°C; Relative humidity: ≤95%.
7 TROUBLESHOOTING
ENGLISH
Trouble Possible Reason Solution
The values can not be
displayed normally or
stably.
1) The nger is not properly inserted.
2) The nger is shaking or the patient is
moving.
3) The device is not used in environment
required by the manual.
4) The device works abnormally.
1) Please insert the nger properly and
measure again.
2) Let the patient keep calm.
3) Please use the device in normal envi-
ronment.
4) Please contact the after-sales.
The device can not be
turned on
1) The battery is drained away or almost
drained away.
2) The battery is installed incorrectly.
3) The device’s malfunction.
1) Please change batteries.
2) Please Install the battery again.
3) Please contact the local service center.
The display disappears
suddenly.
1) The device enters into the energy saving
mode.
2) Low battery.
3) The device works abnormally.
1) Normal.
2) Please charge the battery.
3) Please contact the after-sales.
The data can not be
stored.
1) The device is not operated according to
the manual.
2) The device works abnormally.
1) Please operate the device according to
the manual.
2) Please contact the after-sales.
12


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gima OXY 50 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gima OXY 50 in the language / languages: English as an attachment in your email.

The manual is 1.06 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gima OXY 50

Gima OXY 50 User Manual - French - 19 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info