Frequently Asked Questions
In the case of the TS100 Cable Fault Finder, the variance
occurs in only a few cable types and both readings are
within the specified accuracy of the instrument. The
second cause occurs when the cable is coiled, as in a
box or on a spool. The magnetic field caused by the
TDR signal itself couples across to other parts of the
cable and changes the characteristics of the reflections.
Q: Why, on some cables, does the number displayed
jump between 2 or 3 different values?
A: As the TDR signal travels down a cable, it loses some
of its strength. At some point, the noise on the cable
has an amplitude similar to the reduced strength TDR
signal and will influence the measurement results. The
tester’s software filters out many of the noise related
variations in the displayed length, but some variations
do get through.
Q: I accidentally cracked the plastic housing, does this
affect the moisture protection of the components?
A: Not at all. The component protection is provided by
a coating on the components and Printed Circuit Board
(PCB).
However, if sufficient plastic is missing then a possible
shock hazard exists. You should not use the tester until
the plastic is repaired or replaced.
Q: Can this tester measure the length of single
conductor wires like THHN?
A: No. All TS100 measurements must be made on TWO
conductors from the SAME END of a cable.
Q: If I touch the bare metal of the wires or clip leads,
will the measurement be affected?
A: After BOTH clips are connected, measurement
results will ordinarily not be affected if inappropriate
human contact is made with the input connectors.
Under moist conditions, if a large surface area of cable
is in contact with moist skin, some readings may be
affected.
Q: On multi-conductor cables with a short between
two of the conductors, I sometimes read an “open” at
twice the known length of the cable.
A: If the cable has more than two conductors, and a
short exists at the far end between one of the
conductors you are connected to and a conductor you
are not connected to, the displayed length will be the
SUM of the lengths of the conductors joined by the
short. TS100 can only correctly test the two conductors
you are connected to. See “Applications” on page 7 for
multi-conductor cables.
Q: When testing a set of wires that go into a conduit, I
sometimes get a reading of 0 or 1. Why?
A: If there is more than a foot or so of wires that are
physically separated before they enter the close
confinement of the conduit, this will look to the tester
like an open at the start of the cable. Remember that
TS100 reports the FIRST failure that it finds. Try
bringing the two wires of the pair closer together for
the path from TS100 to the entry to the conduit.
Q: When connecting to a 6 foot (2 meter) piece of 50Ω
Coax with the alligator clips, the tester reads 8 feet (3
meters). What’s up?
A: When measuring a low impedance small cable (less
than 15 feet [5 meters]), the clip leads can add up to 2
feet (1 meter) of length. For longer or high impedance
cables, the clip leads have no effect.
Q: How does the tester react to a speaker or a
transformer at the end of a cable?
A: A speaker or a transformer is actually a large coil of
wire. This will usually cause the length reading to be
larger than that of the cable alone. A moderate power
speaker will add 500 feet (150 meters) to the length
reading. Some combinations of speakers and
transformers connected to the cable may prevent the
tester from making a valid reading.
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke networks TS 100 Cable Fault Finder at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke networks TS 100 Cable Fault Finder in the language / languages: English as an attachment in your email.

The manual is 0,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info