775832
80
Zoom out
Zoom in
Previous page
1/98
Next page
190 Series III
Users Manual
74
Independently Floating Isolated Inputs
The test tool has independently floating isolated inputs. You can use the independently floating
isolated inputs to measure signals that are independently floating from each other. Independently
floating isolated inputs offer additional safety and measurement capabilities compared to inputs
with common references or grounds.
Each input section (A, B, C, D, A, B, METER INPUT) has its own signal input and its own
reference input. The reference input of each input section is electrically isolated from the
reference inputs of the other input sections. The isolated input architecture makes the test tool
about as versatile as having four independent instruments. The advantages of having
independently floating isolated inputs are:
Simultaneous measurement of independently floating signals.
Additional safety:
Since the commons are not directly connected, the chance of causing short circuit when
measuring multiple signals is greatly reduced.
When measuring in systems with multiple grounds, the ground currents induced are kept
to a minimum.
Because the references are not connected together inside the test tool, each reference of the
used inputs must be connected to a reference voltage. Independently floating isolated inputs are
still coupled by parasitic capacitance. This can occur between the input references and the
environment, and between the input references mutually. See Figure 38. For this reason, you
should connect the references to a system ground or another stable voltage. If the reference of
an input is connected to a high speed and / or high voltage signal, you should be aware of
parasitic capacitance. See Figure 38, Figure 40, Figure 41, and Figure 42.
Note
The input channels are electrically isolated from the USB port and from the power
adapter input.
80


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke MDA-550-III at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke MDA-550-III in the language / languages: English as an attachment in your email.

The manual is 9.23 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info