Operating the Meter from the Front Panel
Selecting a Function Modifier 3
3-17
An external TTL signal on pin 9 will trigger a measurement cycle. Alternatively, pin 9 of
the RS-232 interface can be connected to pin 1 through an external switch. See Figure 3-
12. A measurement cycle is trigger when the switch is closed and the +5 volts from pin 1
is applied to pin 9. The trigger event occurs on the rising edge of the signal applied to
pin 9.
Table 3-4. RS-232 Pin Out
1234
6789
5
eue23.eps
Pin # Description Pin # Description
1 +5 V OUT 2 RS-232 RXD
3 RS-232 TXD 5 RS-232 GND
6 Trigger Out 9 Trigger In
Figure 3-12 shows a method for using the +5 V OUT (pin 1) signal with an external
switch to trigger the Meter.
+5 V Out (Pin 1)
Trig In (Pin 9)
eue24.eps
Figure 3-12. External Trigger Circuit
Selecting a Function Modifier
This section describes the function modifiers available with the Meter. Function
modifiers are actions that the Meter performs on an input before a reading is displayed
(for example, a comparison to another value). Function modifiers can be used in
combination. See the “Using Function Modifiers in Combination” section later in this
chapter.
To use a function modifier, press a measurement function button to select that function,
and then press the function modifier button to modify that function. (For example, press
D to select dc voltage measurement, and then press I to select the Touch Hold
function to hold the results of your measurement). Note that modified readings are shown
on the primary display only.
51


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info