Documenting Process Calibrator
Process Instrument Calibration
73
17. Push the Done softkey to accept the calibration
parameters. The display changes to the screen shown
in Figure 32.
gks46s.bmp
Figure 32. Measure and Source Screen for Calibration
18. At this time you can do an automatic test or step
through the test points manually. Push the Auto Test
softkey to have the Product go through the tests
automatically. If necessary, push Abort to exit from the
calibration procedure. The tests start at the first test
point, sources the correct temperature and measures
the corresponding current from the transmitter.
When a measurement is stable and recorded, the
Product goes to the subsequent step. Because the
Product waits until the measurement becomes stable,
the Auto Test works as necessary for instruments with
built-in damping. The error of the expected measured
value is shown in the top left of the measure window.
19. The Product moves to the remaining set of points. For
temperature and electrical parameter calibration, the
points are done automatically. If your source pressure,
the Product stops at each step to let you adjust the
pressure source. When the tests are complete, an error
summary table similar to Figure 33 is shown.
gks47s.bmp
Figure 33. Error Summary Screen
85


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info