764336
57
Zoom out
Zoom in
Previous page
1/130
Next page
Documenting Process Calibrator
Source Mode
45
Damping Measurements
The Product normally applies a software filter to dampen
measurements in all functions except continuity. The
specifications assume that damping is turned on. The
damping method is a running average of the last eight
measurements. Fluke recommends leaving damping on.
Turning damping off may be useful when measurement
response is more important than accuracy or noise
reduction. To turn damping off, push the More Choices
softkey twice, then push the Dampen softkey so that Off is
shown. Push Dampen again to turn damping back on. The
default state is On.
Note
If a measurement falls outside a random noise window,
a new average is started. If damping is turned off, or
until measurements are fully damped, the symbol is
displayed.
Source Mode
The operating mode (for example, MEASURE, SOURCE) is
shown on the display. If the Product is not in SOURCE
mode, push until SOURCE is shown. The Product
must be in SOURCE mode to change any of the SOURCE
parameters.
Source Electrical Parameters
To select an electrical source function:
1. Connect the test leads as shown in Figure 18,
depending on the source function.
2. Push for current, for dc voltage, h for
frequency, or q for resistance.
3. Record the necessary output value, then push . For
example, to source 5.5 V dc, push 5 . .
57


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info