Pressure Calibrator with Electric Pump
Using the Internal Pump
11
To use the Calibrator’s internal pump, refer to Figure 3
and perform the following steps:
1. Depressurize and drain the line before connecting the
Calibrator.
2. Connect the pressure transmitter to the Calibrator
internal sensor as shown in Figure 4 (for internal
pressure sensor measurements) or Figure 5 (for
pressure module measurements).
Note
To avoid leaks, use Teflon tape or similar sealant
on all pressure connections.
3. Make sure the pressure/vacuum switch is in the
desired position. Forward (clockwise) is for pressure;
backward (counter-clockwise) is for vacuum.
4. Turn the pressure/vacuum release control backward
(counter-clockwise) to vent pressure/vacuum from the
pump.
5. Press Zto zero the pressure display.
6. Turn the fine adjustment knob to mid-range.
7. Turn the pressure/vacuum release control forward
(clockwise) to close the release valve.
8. Press
P to apply pressure/vacuum.
Note
This knob adjusts a small internal reservoir to
vary the total volume. With larger external
pressure/vacuum volumes, this control will adjust
pressure or vacuum within a smaller range.
9. Depressurize the system before disconnecting the
pressure line.
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 719 Series at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 719 Series in the language / languages: English as an attachment in your email.

The manual is 1,18 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 719 Series

Fluke 719 Series User Manual - German - 36 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info