Pressure Calibrator with Electric Pump
Zeroing with Absolute Pressure Modules
9
3. Press S to enter pressure switch test mode. The
Calibrator will display CLOSE instead of a mA
measurement.
4. Apply pressure with the pump slowly until the switch
opens.
Note
In the switch test mode, the display update rate is
increased to help capture changing pressure
inputs. Even with the enhanced sample rate,
pressuring the device under test should be done
slowly to ensure accurate readings.
5. OPEN is displayed once the switch is open. Bleed the
pump slowly until the pressure switch closes. RCL
appears on the display.
6. Press S to read the pressure values for when the
switch opened, for when it closed, and for the
deadband
Hold S for 3 seconds to reset Switch Test mode; press
any other key to exit.
Zeroing with Absolute Pressure Modules
For zeroing, adjust the Calibrator to read a known
pressure. This can be barometric pressure, if it is
accurately known, for all but the 700PA3 module. An
accurate pressure standard can also apply a pressure
within range for any Absolute Pressure Module. Adjust the
Calibrator reading as follows:
1. Press and hold Z.
2. Press
Z to increase or Y to decrease the Calibrator
reading to equal the applied pressure.
3. Release Z to exit the zeroing procedure.
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 719 Series at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 719 Series in the language / languages: English as an attachment in your email.

The manual is 1,18 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 719 Series

Fluke 719 Series User Manual - German - 36 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info