Precision Loop Calibrator
Valve Tests (710)
35
Signature Test
In the Signature Test, the product sources a ramping 4 mA to
20 mA to 4 mA signal and records the pressure applied to the
actuator from the positioner and the position.
To set up:
1. Go to Signature Test Menu. See Figure 52.
Figure 52. Signature Test Setup
2. Push the knob to move the cursor through the options.
3. Confirm the Ramp Time, change if needed.
4. Change the Tolerance as required from 1 % to 20 %.
5. In the Calibrator Setup Menu, select Test Tolerance > Valve
Test Parameters to set the default test tolerances.
6. Change the size of steps as required from 0.1 mA to 4.0 mA.
To run the test:
1. Push to start the test.
The Product varies the 4 to 20 mA signal, monitors the TV and
QV variables (pressure and position), and compares it to
expected values.
Good test result: the position variable is as expected.
Marginal test result: the position variable value is found to
be close to the test tolerance.
Bad test result: the position variable deviation values vary
greater than the tolerance.
2. Select Save to save the test results and upload later to the
ValveTrack software.
Or, select Discard if you do not want to save the test result.
Manual Test
Select Manual Test to see the mA signal applied and HART
variables. See Figure 53.
Figure 53. Manual Test
The HART variables that show on the display are important
information about the operation of the valve.
HART variable mapping:
PV: Digitized mA value that represents the mA value applied to the
valve.
Note
This value should agree closely with the applied mA
value.
SV: Setpoint where the valve is expected to be at in % of travel.
Valves forward acting are configured to Fail Open (FO) as the 4 to
20 mA signal moves the valve from 0 % to 100 % of travel (from
open to closed) with 4 to 20 mA signals applied. If the valve is
configured as FC and is forward acting, the valve opens as the
applied mA signal increases to 20 mA.
41


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 709 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 709 in the language / languages: English as an attachment in your email.

The manual is 5,6 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 709

Fluke 709 User Manual - German - 50 pages

Fluke 709 User Manual - Dutch - 50 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info