764223
27
Zoom out
Zoom in
Previous page
1/64
Next page
Digital Multimeters
How to Make Measurements
17
XW Warning
To avoid possible electric shock or personal
injury, do not use the low-pass filter to verify
the presence of hazardous voltages.
Voltages greater than what is indicated may
be present. First, make a voltage
measurement without the filter to detect the
possible presence of hazardous voltage.
Then, select the filter.
Note
When the low-pass filter is selected, the Meter
goes to manual ranging mode. Select ranges by
pressing . Autoranging is not available with
the low-pass filter.
1 kHz
100 Hz
aom11f.eps
Figure 3. Low-Pass Filter
Temperature Measurements (28 II)
The Meter measures the temperature of a type-K
thermocouple (included). Choose between degrees
Celsius (°C) or degrees Fahrenheit (°F) by pushing
.
W Caution
To avoid possible damage to the Meter or
other equipment, remember that while the
Meter is rated for -200.0 °C to +1090.0 °C and
-328.0 °F to 1994 °F, the included type-K
thermocouple is rated to 260 °C. For
temperatures out of that range, use a higher
rated thermocouple.
Display ranges are -200.0 °C to +1090 °C and -328.0 °F
to 1994 °F. Readings outside of these ranges show 0L on
the Meter display. When there is no thermocouple
connected, the display also reads 0PEn.
To measure temperature, do the following:
1. Connect a type-K thermocouple to the Meter’s COM
and terminals.
2. Turn the rotary switch to .
3. Push to enter temperature mode.
4. Push to choose Celsius or Fahrenheit.
27


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 27 II at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 27 II in the language / languages: English as an attachment in your email.

The manual is 1,77 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 27 II

Fluke 27 II User Manual - German - 64 pages

Fluke 27 II User Manual - Dutch - 64 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info