When measuring voltage, the Meter acts approximately
like a 10-MΩ (10,000,000 Ω) impedance in parallel with
the circuit. This loading effect can cause measurement
errors in high-impedance circuits. In most cases, the error
is negligible (0.1 % or less) if the circuit impedance is
10 kΩ (10,000 Ω) or less.
For better accuracy when measuring the dc offset of an
ac voltage, measure the ac voltage first. Note the ac
voltage range, then manually select a dc voltage range
equal to or higher than the ac range. This procedure
improves the accuracy of the dc measurement by
ensuring that the input protection circuits are not
activated.
Zero Input Behavior of True-rms Meters (28 II)
True-rms meters accurately measure distorted
waveforms, but when the input leads are shorted together
in the ac functions, the Meter displays a residual reading
between 1 and 30 counts. When the test leads are open,
the display readings may fluctuate due to interference.
These offset readings are normal. They do not affect the
Meter’s ac measurement accuracy over the specified
measurement ranges.
Unspecified input levels are:
AC voltage: below 3 % of 600 mV ac, or 18 mV ac
AC current: below 3 % of 60 mA ac, or 1.8 mA ac
AC current: below 3 % of 600 μA ac, or 18 μA ac
Low-Pass Filter (28 II)
The 28 II is equipped with an ac low-pass filter. When
measuring ac voltage or ac frequency, press to
activate the low-pass filter mode (). The Meter
continues measuring in the chosen mode, but now the
signal diverts through a filter that blocks unwanted
voltages above 1 kHz, refer to Figure 3. The lower
frequency voltages pass with reduced accuracy to the
measurement below 1 kHz. The low-pass filter can
improve measurement performance on composite sine
waves that are typically generated by inverters and
variable frequency motor drives.
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 27 II at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 27 II in the language / languages: English as an attachment in your email.

The manual is 1,77 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 27 II

Fluke 27 II User Manual - German - 64 pages

Fluke 27 II User Manual - Dutch - 64 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info