45
Inductance
Some inductors are intended to operate at a certain DC bias to achieve a certain
inductance value.However,the LCR meters cannot produce such biasing scheme
and external biasing should not be attempted because external power would be
applied to the instrument and cause serious damage to the meter. Therefore, in
some cases, measured inductance reading may not agree with manufacturer’s
specification. It is important to check if specification pertains to DC biasing or not.
Resistance
When measuring resistance of devices, it is important to know that there are two
types or ways of measurement. One type is DC resistance measurement. Another
type is AC resistance measurement. The LCR meter provide both of types for
measurement. When measuring a resistive component that is designed to be
measured with DC, readings will be incorrect or inaccurate.Before using the meter
to measure resistance, please verify whether the DUT (device under test) requires
DC or AC resistance measurement method. Depending on the method,results will
vary greatly.
8-4 Guard Terminal
One of the input sockets and terminals is labeled as "GUARD". This terminal does
not have to be used in all instances for the meter to make measurements.But in
some instances, it is very useful. Guard terminal generally serves two purposes. If
user is using test leads, the guard terminal can be used to connect to the shielding
of the test leads. Doing so can be useful when making large resistive component
measurements. For example, when measuring a 10 M resistor with test leads, at
the high range the reading may seem to be unstable as a few digits may
continuously be changing. Having the shield of the test leads connected to the
guard terminal will help stabilize the reading in some instances. Guard terminal is
also used to minimize noise and to help minimize parasitic effects coming from the
component to be measured, thus allowing high precision results.
45


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma 6400 LCR Meter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma 6400 LCR Meter in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 0,81 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info