668440
42
Zoom out
Zoom in
Previous page
1/52
Next page
Elma 5800 side 42
Note: The temperature probe is fitted with a type K mini connector. A mini connector to
banana connector adaptor is supplied for connection to the input banana jacks.
Frequency (Duty cycle) measurements (electronic)
1. Set the rotary function switch to the green Hz/% position.
2. Insert the black lead banana plug into the negative COM jack
and the red test lead banana plug into the positive Hz jack.
3. Touch the test probe tips to the circuit under test.
4. Read the frequency on the display.
5. Press the MODE button to indicate “%”.
6. Read the % duty cycle in the display.
% 4 – 20mA MEASUREMENTS
1. Set up and connect as described for DC mA measurements.
2. Set the rotary function switch to the 4-20mA% position.
3. The meter will display loop current as a % with 0mA=-25%, 4mA=0%,
20mA=100%, and 24mA=125%.
Resistance measurements
a) Set the rotary function switch to the INSULATION position, and Press the RANGE
button to chose one of the voltage which display on the top left corner.
b) Connect two testing lines to the tested
c) Push down and hold the “TEST” button /or press the “ LOCK” keystoke first and then
the “TEST” button, if the tested is electriferous and its voltage ( AC/DC) is over 30V, it will
refuse work and no high-voltage testing occurs, simultaneity, it shows “>30V” on the LCD,
the symbol “ ” flashes, and the buzzer warns frequently. if the tested is diselectriferous
or its voltage is lower than 30V, it will enter into the formal testing process and brings the
high-voltage. on the primary display, the insulation resistance in M is indicated in-phase
with analog bar; on the top right corner display, the tested insulation voltage in V (DC) is
indicated, the symbol “ ” flashes and the buzzer warns frequently
d) Being free from the “TEST” button or pushing down the “TEST” button in the “LOCK “
status can exit from the “LOCK” status and shutoff the high-voltagesynchronously, the
resistance values is indicated in the primary display will be held, and the top right corner
display still be in the status of monitoring the insulation voltage for the tested .
e) Subsequently, discharge the balance insulation voltage of the tested through the inner
switch of the meter.
Turning the function switch or press the EXIT button can exit automatically from testing
status during the process
Power tools and small appliances
This test would also apply to other similar equipment that has a line cord. For double
insulated power tools, the megohmmeter lead shown connected to the housing would be
connected to some metal part lf the tool(e..g chuck,blade).
Note:The switch of the device must be in the “ON”position and the main power should be
disconnected.
42


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma 5800 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma 5800 in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 2,02 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info