668440
39
Zoom out
Zoom in
Previous page
1/52
Next page
Elma 5800 side 39
mV voltage measurements
CAUTION: Do not measure mV voltages if a motor on the circuit is being
switched ON or OFF. Large voltage surges may occur that can damage
the meter.
1. Set the function switch to the green mV position.
2. Press the MODE button to indicate “DC”.or ““AC ”, or in AC range
press EXIT for two seconds and chose ”AC+DC”
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
4. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
5. Read the mV voltage in the display.
DC current measurements
CAUTION: Do not make 20A current measurements for longer than 30
seconds. Exceeding 30 seconds may cause damage to the meter and/or
the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA DC, set the function
switch to the yellow µA position and insert the red test lead banana
plug into the µA/mA jack.
3. For current measurements up to 400mA DC, set the function switch
to the yellow mA position and insert the red test lead banana plug
into the µA/mA jack.
4. For current measurements up to 20A DC, set the function switch to
the yellow 10A/HZ/% position and insert the red test lead banana
plug into the 10A jack.
5. Press the MODE button to indicate “DC” on the display.
6. Remove power from the circuit under test, then open up the circuit at the point where
you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Elma 5800 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Elma 5800 in the language / languages: English, Danish, Swedish, Norwegian as an attachment in your email.

The manual is 2,02 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info