801304
17
Zoom out
Zoom in
Previous page
1/44
Next page
Instructions for use POWERSTROLL
17
Drive Medical GmbH & Co. KG | As per: 04.05.2021 | Errors and alterations excepted
EN
Disassembly and transport
1. Switch o the PowerStroll and unplug the cable at
the control panel.
2. Remove the relief strap from the push handle.
3. To release the battery, press the gray PUSH but-
ton on the power unit (photo 16). Then pull out the
battery.
4. Remove the power unit. To do this, pull the black
band between the holders upwards and at the same
time, take the power unit out of the holder from the
back (see image below).
5. Stow all parts in the carrying bag provided.
Note
The battery supplied with the PowerStroll meets IATA
special provision A67 and is approved for transport in an
airplane. Nonetheless, if you have any questions in this
regard, consult your airline prior to travel.
Cleaning | Maintenance
Clean regularly with a moist cloth!
Do not use any aggressive cleaning agents!
Do not use abrasive cleaners!
Do not use any high-dose chlorine and vinegar-based
cleaners!
To disinfect, a spray disinfectant can be used such as
Descosept AF or Bacillol AF.
Check the screws regularly and tighten, if necessary.
The product must not be continuously exposed to tem-
peratures over 60°C.
In the event of defects, contact your specialist dealer
from whom you purchased this product.
Do not use any petroleum ether, solutions or evapora-
ting solutions since these could damage plastic parts.
Do not use any wax.
Ensure that the charging unit was unplugged and that
the PowerStroll is switched o before you clean it.
Check the condition of the tires regularly. They are
"fail-safe," meaning that you do not have to check air
pressure or fill them with air. Always check the tires for
tears or other damage, however.
Troubleshooting
The wheelchair does not move although the lever is
squeezed:
Check whether the PowerStroll is switched on.
Check whether the battery is fully charged.
Check whether the connection between the battery
and motor is correct.
Check whether the brakes of the wheelchair are
released.
Check whether the drive wheel is in contact with the
ground.
It is dicult to move the wheelchair by hand:
Check whether the brakes of the wheelchair are
released.
• Raise the PowerStroll.
Warranty conditions
There is a comprehensive warranty for a period of two
years (except for parts subject to wear; 6 months for the
battery) starting from the date of delivery of your new Po-
werStroll. The warranty includes the replacement parts
of the PowerStroll during this period.
Technical data
PowerStroll PowerStroll HD
Article number PWCPP011 PWCPP011HD
HMV (medical
aids register) no. 18.99.04.1024
Overall width 25 cm 25 cm
Overall length 41 cm 41 cm
Speed 6 km/h 6 km/h
Battery
capacity 2x 12 Ah 12 Ah
Motor output 150 W 200 W
Climbing ability 10,5 % 10,5 %
Total weight 19,9 kg 19,9 kg
Weight of battery 9,7 kg 9,7 kg
Weight without
battery 10,2 kg 10,2 kg
Max. load 135 kg 180 kg
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Drive PowerStroll at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Drive PowerStroll in the language / languages: English, German, Dutch, French, Italian as an attachment in your email.

The manual is 4.28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info