Use and Operation
IImmppoorrttaanntt aaddvviissee ffoorr ssttoorraaggee
The goods should only be stored in the provided
drawers and/or racks.
The loading must never obstruct the air-inlet-
openings inside the unit.
Make sure that the loaded goods do not tilt to the
backside.
Make sure that the loading does not touch the
reference bottle of the display sensor.
Do not overload the unit.
AAllaarrmm ssiittuuaattiioonnss
On the occurrence of an alarm situation attempts
must be made to find out the reason for the
alarm and remove it as quickly as possible. If that
is not successful the necessary measures must
immediately be taken so that the stored goods do
not sustain any damage.
At a temperature alarm the red LED alarm lights
up and an acoustic signal sounds. In addition the
corresponding alarm message appears on the
display.
In the case of a powerfail alarm the green power
LED and the red LED alarm flash and an acoustic
signal sounds. The display is switched off but be
temporarly switched on again by pressing any
key.
After a given time (setting: EXTRA - DISPLAY -
BACKL. TIMEOUT ) the display switches off again.
The acoustic alarm can be suppressed for a set
time by pressing the ENTER button. During a
powerfail phase no parameters can be changed.
If the internal temperature rises above the upper
limit during a powerfail phase a temperature
alarm is triggered and entered into the History
list. In the case of a door alarm the red LED lights
up and an acoustic signal sounds. In addition the
corresponding alarm message appears on the
display
AAllaarrmm HHiissttoorryy
The alarm history list contains all relevant data
on temperature and powerfail events. The list
contains up to 20 alarm situations.
The status message “HISTORY” indicates that
new entries have been made in the list.
The alarm history list can be accessed via the
“SHOW HISTORY” function in the EXTRA menu.
Within the list you navigate by means of the UP
and DOWN buttons. The first and last entry in the
list are marked “BEGIN OF LIST” and “END OF
LIST”
The following information is stored in the Alarm History :
Temperature alarm :
ALARM : type of alarm (HIGH ALARM or LOW ALARM)
START : Start date and time of alarm situation
END : End date and time of alarm situation
AVG : Average temperature on the display sensor during the alarm situations.
MAX / MIN : Maximum or minimum temperature value on the display sensor during
the
alarm situation (depending on the type of temperature alarm)
Powerfail alarm :
ALARM : POWERFAIL
START : Start date and time of alarm situation
END : End date and time of alarm situation
AVG : Average temperature on the display sensor during the alarm situations
To delete the alarm history, select the function “DELETE HISTORY” from the EXTRA menu.
12


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Dometic ML155 SG at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Dometic ML155 SG in the language / languages: English as an attachment in your email.

The manual is 0,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Dometic ML155 SG

Dometic ML155 SG User Manual - German - 14 pages

Dometic ML155 SG User Manual - Dutch - 14 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info